Abstract: Photon-counting Computed Tomography (PCCT) is a promising imaging technique. Using detectors that count the number and energy of photons in multiple bins, PCCT offers several advantages over conventional CT, including a higher image quality, reduced contrast agent volume, radiation doses, and artifacts. Although PCCT is well established for cardiac imaging in assessing coronary artery disease, its application in aortic imaging remains limited. This review summarizes the available literature and provides an overview of the current use of PCCT for the diagnosis of aortic imaging, focusing mainly on endoleaks detection and characterization after endovascular aneurysm repair (EVAR), contrast dose volume, and radiation exposure reduction, particularly in patients with chronic kidney disease and in those requiring follow-up CT.

Advantages of photon-counting detector CT in aortic imaging / Zanon, Chiara; Cademartiri, Filippo; Toniolo, Alessandro; Bini, Costanza; Clemente, Alberto; Chiara Colacchio, Elda; Cabrelle, Giulio; Mastro, Florinda; Antonello, Michele; Quaia, Emilio; Pepe, Alessia. - In: TOMOGRAPHY. - ISSN 2379-139X. - 10:1(2023), pp. 1-13. [10.3390/tomography10010001]

Advantages of photon-counting detector CT in aortic imaging

Chiara Zanon;Alessia Pepe
2023

Abstract

Abstract: Photon-counting Computed Tomography (PCCT) is a promising imaging technique. Using detectors that count the number and energy of photons in multiple bins, PCCT offers several advantages over conventional CT, including a higher image quality, reduced contrast agent volume, radiation doses, and artifacts. Although PCCT is well established for cardiac imaging in assessing coronary artery disease, its application in aortic imaging remains limited. This review summarizes the available literature and provides an overview of the current use of PCCT for the diagnosis of aortic imaging, focusing mainly on endoleaks detection and characterization after endovascular aneurysm repair (EVAR), contrast dose volume, and radiation exposure reduction, particularly in patients with chronic kidney disease and in those requiring follow-up CT.
2023
aortic imaging; vascular; photon-counting; CT angiography; dose exposure; contrast agents; EVAR; endoleaks
01 Pubblicazione su rivista::01g Articolo di rassegna (Review)
Advantages of photon-counting detector CT in aortic imaging / Zanon, Chiara; Cademartiri, Filippo; Toniolo, Alessandro; Bini, Costanza; Clemente, Alberto; Chiara Colacchio, Elda; Cabrelle, Giulio; Mastro, Florinda; Antonello, Michele; Quaia, Emilio; Pepe, Alessia. - In: TOMOGRAPHY. - ISSN 2379-139X. - 10:1(2023), pp. 1-13. [10.3390/tomography10010001]
File allegati a questo prodotto
File Dimensione Formato  
Zanon_Advantages-of-Photon-Counting_2023.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 8.83 MB
Formato Adobe PDF
8.83 MB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1699210
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 2
  • ???jsp.display-item.citation.isi??? ND
social impact