A new low-energy X-ray detector was built and operated, using a plastic scintillator coupled to a large area SiPM. The signal is amplified with a low-noise, high gain, custom circuit providing excellent photon-counting capabilities and allowing a quasi-digital measurement. The detector was tested using X-rays coming from molybdenum K lines (17.4 and 19.6 keV), and an energy resolution of 28% is obtained with 20 photoelectrons per X-ray photon on average.
X-ray spectroscopy with a photon-counting SiPM-based scintillation detector / Ceravolo, S.; Gargiulo, R.; Paesani, D.; Russo, A.; Sarra, I.. - In: JOURNAL OF INSTRUMENTATION. - ISSN 1748-0221. - 18:9(2023), pp. 1-10. [10.1088/1748-0221/18/09/T09007]
X-ray spectroscopy with a photon-counting SiPM-based scintillation detector
Gargiulo, R.
;
2023
Abstract
A new low-energy X-ray detector was built and operated, using a plastic scintillator coupled to a large area SiPM. The signal is amplified with a low-noise, high gain, custom circuit providing excellent photon-counting capabilities and allowing a quasi-digital measurement. The detector was tested using X-rays coming from molybdenum K lines (17.4 and 19.6 keV), and an energy resolution of 28% is obtained with 20 photoelectrons per X-ray photon on average.File | Dimensione | Formato | |
---|---|---|---|