OBJECTIVES This study compared the diagnostic performance with adenosine-derived fractional flow reserve (FFR) <= 0.8 of contrast-based FFR (cFFR), resting distal pressure (Pd)/aortic pressure (Pa), and the instantaneous wave-free ratio (iFR). BACKGROUND FFR objectively identifies lesions that benefit from medical therapy versus revascularization. However, FFR requires maximal vasodilation, usually achieved with adenosine. Radiographic contrast injection causes submaximal coronary hyperemia. Therefore, intracoronary contrast could provide an easy and inexpensive tool for predicting FFR. METHODS We recruited patients undergoing routine FFR assessment and made paired, repeated measurements of all physiology metrics (Pd/Pa, iFR, cFFR, and FFR). Contrast medium and dose were per local practice, as was the dose of intracoronary adenosine. Operators were encouraged to perform both intracoronary and intravenous adenosine assessments and a final drift check to assess wire calibration. A central core lab analyzed blinded pressure tracings in a standardized fashion. RESULTS A total of 763 subjects were enrolled from 12 international centers. Contrast volume was 8 +/- 2 ml per measurement, and 8 different contrast media were used. Repeated measurements of each metric showed a bias < 0.005, but a lower SD (less variability) for cFFR than resting indexes. Although Pd/Pa and iFR demonstrated equivalent performance against FFR <= 0.8 (78.5% vs. 79.9% accuracy; p = 0.78; area under the receiver-operating characteristic curve: 0.875 vs. 0.881; p = 0.35), cFFR improved both metrics (85.8% accuracy and 0.930 area; p < 0.001 for each) with an optimal binary threshold of 0.83. A hybrid decision-making strategy using cFFR required adenosine less often than when based on either Pd/Pa or iFR. CONCLUSIONS cFFR provides diagnostic performance superior to that of Pd/Pa or iFR for predicting FFR. For clinical scenarios or health care systems in which adenosine is contraindicated or prohibitively expensive, cFFR offers a universal technique to simplify invasive coronary physiological assessments. Yet FFR remains the reference standard for diagnostic certainty as even cFFR reached only similar to 85% agreement. (C) 2016 by the American College of Cardiology Foundation.

Continuum of Vasodilator Stress From Rest to Contrast Medium to Adenosine Hyperemia for Fractional Flow Reserve Assessment / Johnson, Nils P; Jeremias, Allen; Zimmermann, Frederik M; Adjedj, Julien; Witt, Nils; Hennigan, Barry; Koo, Bon Kwon; Maehara, Akiko; Matsumura, Mitsuaki; Barbato, Emanuele; Esposito, Giovanni; Trimarco, Bruno; Rioufol, Gilles; Park, Seung Jung; Yang, Hyoung Mo; Baptista, Sérgio B; Chrysant, George S; Leone, Antonio M; Berry, Colin; De Bruyne, Bernard; Gould, K. Lance; Kirkeeide, Richard L; Oldroyd, Keith G; Pijls, Nico H. J; Fearon, William F.. - In: JACC: CARDIOVASCULAR INTERVENTIONS. - ISSN 1936-8798. - 9:8(2016), pp. 757-767. [10.1016/j.jcin.2015.12.273]

Continuum of Vasodilator Stress From Rest to Contrast Medium to Adenosine Hyperemia for Fractional Flow Reserve Assessment

BARBATO, EMANUELE;
2016

Abstract

OBJECTIVES This study compared the diagnostic performance with adenosine-derived fractional flow reserve (FFR) <= 0.8 of contrast-based FFR (cFFR), resting distal pressure (Pd)/aortic pressure (Pa), and the instantaneous wave-free ratio (iFR). BACKGROUND FFR objectively identifies lesions that benefit from medical therapy versus revascularization. However, FFR requires maximal vasodilation, usually achieved with adenosine. Radiographic contrast injection causes submaximal coronary hyperemia. Therefore, intracoronary contrast could provide an easy and inexpensive tool for predicting FFR. METHODS We recruited patients undergoing routine FFR assessment and made paired, repeated measurements of all physiology metrics (Pd/Pa, iFR, cFFR, and FFR). Contrast medium and dose were per local practice, as was the dose of intracoronary adenosine. Operators were encouraged to perform both intracoronary and intravenous adenosine assessments and a final drift check to assess wire calibration. A central core lab analyzed blinded pressure tracings in a standardized fashion. RESULTS A total of 763 subjects were enrolled from 12 international centers. Contrast volume was 8 +/- 2 ml per measurement, and 8 different contrast media were used. Repeated measurements of each metric showed a bias < 0.005, but a lower SD (less variability) for cFFR than resting indexes. Although Pd/Pa and iFR demonstrated equivalent performance against FFR <= 0.8 (78.5% vs. 79.9% accuracy; p = 0.78; area under the receiver-operating characteristic curve: 0.875 vs. 0.881; p = 0.35), cFFR improved both metrics (85.8% accuracy and 0.930 area; p < 0.001 for each) with an optimal binary threshold of 0.83. A hybrid decision-making strategy using cFFR required adenosine less often than when based on either Pd/Pa or iFR. CONCLUSIONS cFFR provides diagnostic performance superior to that of Pd/Pa or iFR for predicting FFR. For clinical scenarios or health care systems in which adenosine is contraindicated or prohibitively expensive, cFFR offers a universal technique to simplify invasive coronary physiological assessments. Yet FFR remains the reference standard for diagnostic certainty as even cFFR reached only similar to 85% agreement. (C) 2016 by the American College of Cardiology Foundation.
2016
adenosine; contrast medium; fractional flow reserve; hyperemia; instantaneous wave-free ratio
01 Pubblicazione su rivista::01a Articolo in rivista
Continuum of Vasodilator Stress From Rest to Contrast Medium to Adenosine Hyperemia for Fractional Flow Reserve Assessment / Johnson, Nils P; Jeremias, Allen; Zimmermann, Frederik M; Adjedj, Julien; Witt, Nils; Hennigan, Barry; Koo, Bon Kwon; Maehara, Akiko; Matsumura, Mitsuaki; Barbato, Emanuele; Esposito, Giovanni; Trimarco, Bruno; Rioufol, Gilles; Park, Seung Jung; Yang, Hyoung Mo; Baptista, Sérgio B; Chrysant, George S; Leone, Antonio M; Berry, Colin; De Bruyne, Bernard; Gould, K. Lance; Kirkeeide, Richard L; Oldroyd, Keith G; Pijls, Nico H. J; Fearon, William F.. - In: JACC: CARDIOVASCULAR INTERVENTIONS. - ISSN 1936-8798. - 9:8(2016), pp. 757-767. [10.1016/j.jcin.2015.12.273]
File allegati a questo prodotto
File Dimensione Formato  
Nils_Continuum-of-Vasodilator-Stress_2016.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 1.38 MB
Formato Adobe PDF
1.38 MB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1660302
Citazioni
  • ???jsp.display-item.citation.pmc??? 34
  • Scopus 124
  • ???jsp.display-item.citation.isi??? 117
social impact