Purpose: We investigated whether optical coherence tomography (OCT)-guided bioresorbable vascular scaffolds (BVS) implantation can improve in-scaffold minimal lumen area (MLA) at 6-month compared with angiography guidance. Methods: The OPTICO BVS was a randomized, international multicenter, assessor blind, superiority trial comparing OCT- versus angiography-guided percutaneous coronary intervention (PCI) (1:1 allocation) in patients with coronary artery disease undergoing Absorb BVS 1.1 implantation. The primary endpoint was in-scaffold MLA at 6-month. Results: The trial was prematurely stopped on May 31, 2017 after enrollment of 38 of 270 planned patients (14%) following the retraction of the device in Europe. Patients were randomly assigned to OCT- (n = 19) or angiography-guided PCI (n = 19). Scaffold diameter (OCT 3.0 ± 0.3 mm vs. angiography 3.1 ± 0.3 mm, P =. 333) and length (28.8 ± 13.6 mm vs. 23.8 ± 12.3 mm, P =. 223) were comparable. There was no significant difference in in-scaffold MLA at 6 months (4.47mm2 vs. 5.08mm2, P =. 692). Scaffold expansion at 6-month was significantly higher in the OCT-guided PCI as compared with angiography-guided PCI (84.5% vs. 76.5%, P = .010). There was no significant difference in clinical outcomes. Conclusions: Although in-scaffold MLA at 6-month did not differ between groups, scaffold expansion was improved following OCT- as compared with angiography-guided PCI. The findings of this study must be interpreted in view of the premature termination with inclusion of 14% of the initially planned study sample.

Randomized Comparison of Optical Coherence Tomography Versus Angiography to Guide Bioresorbable Vascular Scaffold Implantation. The OPTICO BVS Study / Ueki, Y.; Yamaji, K.; Barbato, E.; Nef, H.; Brugaletta, S.; Alfonso, F.; Hill, J.; Cook, S.; Burzotta, F.; Karagiannis, A.; Windecker, S.; Raber, L.. - In: CARDIOVASCULAR REVASCULARIZATION MEDICINE. - ISSN 1553-8389. - 21:10(2020), pp. 1244-1250. [10.1016/j.carrev.2020.03.023]

Randomized Comparison of Optical Coherence Tomography Versus Angiography to Guide Bioresorbable Vascular Scaffold Implantation. The OPTICO BVS Study

Barbato E.;
2020

Abstract

Purpose: We investigated whether optical coherence tomography (OCT)-guided bioresorbable vascular scaffolds (BVS) implantation can improve in-scaffold minimal lumen area (MLA) at 6-month compared with angiography guidance. Methods: The OPTICO BVS was a randomized, international multicenter, assessor blind, superiority trial comparing OCT- versus angiography-guided percutaneous coronary intervention (PCI) (1:1 allocation) in patients with coronary artery disease undergoing Absorb BVS 1.1 implantation. The primary endpoint was in-scaffold MLA at 6-month. Results: The trial was prematurely stopped on May 31, 2017 after enrollment of 38 of 270 planned patients (14%) following the retraction of the device in Europe. Patients were randomly assigned to OCT- (n = 19) or angiography-guided PCI (n = 19). Scaffold diameter (OCT 3.0 ± 0.3 mm vs. angiography 3.1 ± 0.3 mm, P =. 333) and length (28.8 ± 13.6 mm vs. 23.8 ± 12.3 mm, P =. 223) were comparable. There was no significant difference in in-scaffold MLA at 6 months (4.47mm2 vs. 5.08mm2, P =. 692). Scaffold expansion at 6-month was significantly higher in the OCT-guided PCI as compared with angiography-guided PCI (84.5% vs. 76.5%, P = .010). There was no significant difference in clinical outcomes. Conclusions: Although in-scaffold MLA at 6-month did not differ between groups, scaffold expansion was improved following OCT- as compared with angiography-guided PCI. The findings of this study must be interpreted in view of the premature termination with inclusion of 14% of the initially planned study sample.
2020
bioresorbable scaffold; optical coherence tomography; percutaneous coronary intervention
01 Pubblicazione su rivista::01a Articolo in rivista
Randomized Comparison of Optical Coherence Tomography Versus Angiography to Guide Bioresorbable Vascular Scaffold Implantation. The OPTICO BVS Study / Ueki, Y.; Yamaji, K.; Barbato, E.; Nef, H.; Brugaletta, S.; Alfonso, F.; Hill, J.; Cook, S.; Burzotta, F.; Karagiannis, A.; Windecker, S.; Raber, L.. - In: CARDIOVASCULAR REVASCULARIZATION MEDICINE. - ISSN 1553-8389. - 21:10(2020), pp. 1244-1250. [10.1016/j.carrev.2020.03.023]
File allegati a questo prodotto
File Dimensione Formato  
Ueki_Randomized-Comparison-ofOptical-Coherence_2020.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 684.11 kB
Formato Adobe PDF
684.11 kB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1659165
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 7
social impact