Technical improvements and labeling updates of the AFX2 stent graft (Endologix Inc, Irvine, CA) seemed to have solved the known issues of its previous generation (AFX Strata). Although most endograft failures after endovascular abdominal aortic aneurysm repair will be managed endovascularly, a small subset of patients will still require secondary open conversion. Partial or complete endograft removal can be required, mainly dependent on the characteristics of the stent graft previously placed. We have report a case of secondary open conversion for late type Ia/IIIb endoleak due to stent fracture and fabric tear of the AFX2 stent graft 3 years after endovascular abdominal aortic aneurysm repair.
Surgery for late type Ia/IIIb endoleak from a fabric tear and stent fracture of AFX2 stent graft / Cuozzo, S.; Miceli, F.; Marzano, A.; Martinelli, O.; Gattuso, R.; Sbarigia, E.. - In: JOURNAL OF VASCULAR SURGERY CASES AND INNOVATIVE TECHNIQUES. - ISSN 2468-4287. - 8:3(2022), pp. 458-461. [10.1016/j.jvscit.2022.06.005]
Surgery for late type Ia/IIIb endoleak from a fabric tear and stent fracture of AFX2 stent graft
Cuozzo S.
;Miceli F.;Marzano A.;Martinelli O.;Gattuso R.;Sbarigia E.
2022
Abstract
Technical improvements and labeling updates of the AFX2 stent graft (Endologix Inc, Irvine, CA) seemed to have solved the known issues of its previous generation (AFX Strata). Although most endograft failures after endovascular abdominal aortic aneurysm repair will be managed endovascularly, a small subset of patients will still require secondary open conversion. Partial or complete endograft removal can be required, mainly dependent on the characteristics of the stent graft previously placed. We have report a case of secondary open conversion for late type Ia/IIIb endoleak due to stent fracture and fabric tear of the AFX2 stent graft 3 years after endovascular abdominal aortic aneurysm repair.File | Dimensione | Formato | |
---|---|---|---|
Cuozzo_Surgery_2022.pdf
solo gestori archivio
Tipologia:
Documento in Post-print (versione successiva alla peer review e accettata per la pubblicazione)
Licenza:
Tutti i diritti riservati (All rights reserved)
Dimensione
816.31 kB
Formato
Adobe PDF
|
816.31 kB | Adobe PDF | Contatta l'autore |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.