This study presents the initial experimental validation of a mono-static microwave imaging system to monitor liver microwave thermal ablation. The system consists of a tank filled with a coupling medium in which a 3D-printed phantom mimicking the ablation region is embedded. A slotloaded Vivaldi antenna is moved in front of the phantom along a linear path, measuring the S-parameters on a finite number of positions. Both simulation and measurement results are found to agree each other. The differential signal of the microwave imaging system with and without the presence of the phantom is over -90 dB, which is above the detection capacity of commercial VNA. The experimental validation of the system paves the way for the design of a multi-static microwave imaging system.

Initial Experimental Validation of a Microwave Imaging System to Monitor Liver Microwave Thermal Ablation / Wang, Mengchu; Cavagnaro, Marta; Abedi, Soroush; Joachimowicz, Nadine; Roussel, Hélène; Costanzo, Sandra; Crocco, Lorenzo. - (2022). (Intervento presentato al convegno 16th European Conference on Antennas and Propagation tenutosi a Madrid, Spagna).

Initial Experimental Validation of a Microwave Imaging System to Monitor Liver Microwave Thermal Ablation

Mengchu Wang;Marta Cavagnaro;
2022

Abstract

This study presents the initial experimental validation of a mono-static microwave imaging system to monitor liver microwave thermal ablation. The system consists of a tank filled with a coupling medium in which a 3D-printed phantom mimicking the ablation region is embedded. A slotloaded Vivaldi antenna is moved in front of the phantom along a linear path, measuring the S-parameters on a finite number of positions. Both simulation and measurement results are found to agree each other. The differential signal of the microwave imaging system with and without the presence of the phantom is over -90 dB, which is above the detection capacity of commercial VNA. The experimental validation of the system paves the way for the design of a multi-static microwave imaging system.
2022
16th European Conference on Antennas and Propagation
04 Pubblicazione in atti di convegno::04d Abstract in atti di convegno
Initial Experimental Validation of a Microwave Imaging System to Monitor Liver Microwave Thermal Ablation / Wang, Mengchu; Cavagnaro, Marta; Abedi, Soroush; Joachimowicz, Nadine; Roussel, Hélène; Costanzo, Sandra; Crocco, Lorenzo. - (2022). (Intervento presentato al convegno 16th European Conference on Antennas and Propagation tenutosi a Madrid, Spagna).
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1654090
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact