Aims: We report the 6-year results of our use of the Nellix endovascular aneurysm sealing system. Materials and methods: This is a retrospective review of patients we treated from 2013 to 2019. The outcomes described include technical success rate, mortality, all procedure-related complications, reinterventions, open surgical conversion and secondary sac rupture. Results and conclusions: A total of 68 patients were treated. Mortality was 2.9%. Secondary interventions were carried out for distal embolization (3), stent occlusion (1), pseudoaneurysm (1) and endoleak (5). Open surgical conversion was required in 6.4% of cases. Medium-and long-Term complications of EVAS occurred more frequently than expected. Because the durability of endovascular aneurysm sealing is questionable, strict postoperative surveillance of Nellix is crucial to identify features of failure.

Nellix endovascular aneurysm-sealing system. A single-center experience and review of current evidence / Martinelli, O.; Alunno, A.; Gattuso, R.; Di Girolamo, A.; Irace, L.. - In: FUTURE CARDIOLOGY. - ISSN 1479-6678. - 17:5(2021), pp. 875-884. [10.2217/fca-2020-0134]

Nellix endovascular aneurysm-sealing system. A single-center experience and review of current evidence

Martinelli O.;Alunno A.
;
Gattuso R.;Di Girolamo A.;Irace L.
2021

Abstract

Aims: We report the 6-year results of our use of the Nellix endovascular aneurysm sealing system. Materials and methods: This is a retrospective review of patients we treated from 2013 to 2019. The outcomes described include technical success rate, mortality, all procedure-related complications, reinterventions, open surgical conversion and secondary sac rupture. Results and conclusions: A total of 68 patients were treated. Mortality was 2.9%. Secondary interventions were carried out for distal embolization (3), stent occlusion (1), pseudoaneurysm (1) and endoleak (5). Open surgical conversion was required in 6.4% of cases. Medium-and long-Term complications of EVAS occurred more frequently than expected. Because the durability of endovascular aneurysm sealing is questionable, strict postoperative surveillance of Nellix is crucial to identify features of failure.
2021
abdominal aortic aneurysm; endoleak; endovascular sealing; EVAS; migration; Nellix; blood vessel prosthesis; humans; prosthesis design; retrospective studies; stents; treatment outcome; aortic aneurysm, abdominal; blood vessel prosthesis implantation; endovascular procedures
01 Pubblicazione su rivista::01a Articolo in rivista
Nellix endovascular aneurysm-sealing system. A single-center experience and review of current evidence / Martinelli, O.; Alunno, A.; Gattuso, R.; Di Girolamo, A.; Irace, L.. - In: FUTURE CARDIOLOGY. - ISSN 1479-6678. - 17:5(2021), pp. 875-884. [10.2217/fca-2020-0134]
File allegati a questo prodotto
File Dimensione Formato  
Martinelli_Nellix-endovascular-aneurysm-sealing_2021.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 5.24 MB
Formato Adobe PDF
5.24 MB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1648050
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact