Objectives To investigate the diagnostic accuracy of CT coronary artery calcium scoring (CACS) with tin pre-filtration (Sn100 kVp) using iterative beam-hardening correction (IBHC) calcium material reconstruction compared to the standard 120 kVp acquisition. Background Third generation dual-source CT (DSCT) CACS with Sn100 kVp acquisition allows significant dose reduction. However, the Sn100 kVp spectrum is harder with lower contrast compared to 120kVp, resulting in lower calcium score values. Sn100 kVp spectral correction using IBHC-based calcium material reconstruction may restore comparable calcium values. Methods Image data of 62 patients (56% male, age 63.9 ± 9.2years) who underwent a clinically-indicated CACS acquisition using the standard 120 kVp protocol and an additional Sn100 kVp CACS scan as part of a research study were retrospectively analyzed. Datasets of the Sn100 kVp scans were reconstructed using a dedicated spectral IBHC CACS reconstruction to restore the spectral response of 120 kVp spectra. Agatston scores were derived from 120 kVp and IBHC reconstructed Sn100 kVp studies. Pearson's correlation coefficient was assessed and Agatston score categories and percentile-based risk categorization were compared. Results Median Agatston scores derived from IBHC Sn100 kVp scans and 120 kVp acquisition were 31.7 and 34.1, respectively (p = 0.057). Pearson's correlation coefficient showed excellent correlation between the acquisitions (r = 0.99, p < 0.0001). Agatston score categories and percentile-based cardiac risk categories showed excellent agreement (ĸ = 1.00 and ĸ = 0.99), resulting in a low cardiac risk reclassification of 1.6% with the use of IBHC CACS reconstruction. Image noise was 24.9 ± 3.6HU in IBHC Sn100 kVp and 17.1 ± 3.9HU in 120 kVp scans (p < 0.0001). The dose-length-product was 13.2 ± 3.4 mGy cm with IBHC Sn100 kVp and 59.1 ± 22.9 mGy cm with 120 kVp scans (p < 0.0001), resulting in a significantly lower effective radiation dose (0.19 ± 0.07 mSv vs. 0.83 ± 0.33 mSv, p < 0.0001) for IBHC Sn100 kVp scans. Conclusion Low voltage CACS with tin filtration using a dedicated IBHC CACS material reconstruction algorithm shows excellent correlation and agreement with the standard 120 kVp acquisition regarding Agatston score and cardiac risk categorization, while radiation dose is significantly reduced by 75% to the level of a chest x-ray.

CT coronary calcium scoring with tin filtration using iterative beam-hardening calcium correction reconstruction / Tesche, C.; De Cecco, C. N.; Schoepf, U. J.; Duguay, T. M.; Albrecht, M. H.; De Santis, D.; Varga-Szemes, A.; Lesslie, V. W.; Ebersberger, U.; Bayer, R. R.; Canstein, C.; Hoffmann, E.; Allmendinger, T.; Nance, J. W.. - In: EUROPEAN JOURNAL OF RADIOLOGY. - ISSN 0720-048X. - 91:(2017), pp. 29-34. [10.1016/j.ejrad.2017.03.011]

CT coronary calcium scoring with tin filtration using iterative beam-hardening calcium correction reconstruction

De Cecco C. N.
Secondo
;
De Santis D.;
2017

Abstract

Objectives To investigate the diagnostic accuracy of CT coronary artery calcium scoring (CACS) with tin pre-filtration (Sn100 kVp) using iterative beam-hardening correction (IBHC) calcium material reconstruction compared to the standard 120 kVp acquisition. Background Third generation dual-source CT (DSCT) CACS with Sn100 kVp acquisition allows significant dose reduction. However, the Sn100 kVp spectrum is harder with lower contrast compared to 120kVp, resulting in lower calcium score values. Sn100 kVp spectral correction using IBHC-based calcium material reconstruction may restore comparable calcium values. Methods Image data of 62 patients (56% male, age 63.9 ± 9.2years) who underwent a clinically-indicated CACS acquisition using the standard 120 kVp protocol and an additional Sn100 kVp CACS scan as part of a research study were retrospectively analyzed. Datasets of the Sn100 kVp scans were reconstructed using a dedicated spectral IBHC CACS reconstruction to restore the spectral response of 120 kVp spectra. Agatston scores were derived from 120 kVp and IBHC reconstructed Sn100 kVp studies. Pearson's correlation coefficient was assessed and Agatston score categories and percentile-based risk categorization were compared. Results Median Agatston scores derived from IBHC Sn100 kVp scans and 120 kVp acquisition were 31.7 and 34.1, respectively (p = 0.057). Pearson's correlation coefficient showed excellent correlation between the acquisitions (r = 0.99, p < 0.0001). Agatston score categories and percentile-based cardiac risk categories showed excellent agreement (ĸ = 1.00 and ĸ = 0.99), resulting in a low cardiac risk reclassification of 1.6% with the use of IBHC CACS reconstruction. Image noise was 24.9 ± 3.6HU in IBHC Sn100 kVp and 17.1 ± 3.9HU in 120 kVp scans (p < 0.0001). The dose-length-product was 13.2 ± 3.4 mGy cm with IBHC Sn100 kVp and 59.1 ± 22.9 mGy cm with 120 kVp scans (p < 0.0001), resulting in a significantly lower effective radiation dose (0.19 ± 0.07 mSv vs. 0.83 ± 0.33 mSv, p < 0.0001) for IBHC Sn100 kVp scans. Conclusion Low voltage CACS with tin filtration using a dedicated IBHC CACS material reconstruction algorithm shows excellent correlation and agreement with the standard 120 kVp acquisition regarding Agatston score and cardiac risk categorization, while radiation dose is significantly reduced by 75% to the level of a chest x-ray.
2017
coronary artery calcium score; coronary artery disease; coronary computed tomographic angiography; tin filtration
01 Pubblicazione su rivista::01a Articolo in rivista
CT coronary calcium scoring with tin filtration using iterative beam-hardening calcium correction reconstruction / Tesche, C.; De Cecco, C. N.; Schoepf, U. J.; Duguay, T. M.; Albrecht, M. H.; De Santis, D.; Varga-Szemes, A.; Lesslie, V. W.; Ebersberger, U.; Bayer, R. R.; Canstein, C.; Hoffmann, E.; Allmendinger, T.; Nance, J. W.. - In: EUROPEAN JOURNAL OF RADIOLOGY. - ISSN 0720-048X. - 91:(2017), pp. 29-34. [10.1016/j.ejrad.2017.03.011]
File allegati a questo prodotto
File Dimensione Formato  
Tesche_postprint_CT coronary calcium_2017.pdf

Open Access dal 24/03/2018

Tipologia: Documento in Post-print (versione successiva alla peer review e accettata per la pubblicazione)
Licenza: Creative commons
Dimensione 629.45 kB
Formato Adobe PDF
629.45 kB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1634201
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 17
  • ???jsp.display-item.citation.isi??? 17
social impact