Purpose: To investigate the effect of an iterative beam-hardening correction algorithm (iBHC) on artifact reduction and image quality in coronary CT angiography (cCTA) with low tube voltage. Material and methods: Thirty-six patients (17 male, mean age, 57.3 ± 14.5 years) were prospectively enrolled in this IRB-approved study and underwent 70-kV cCTA using a third-generation dual-source CT scanner. Images were reconstructed using a standard algorithm (Bv36) both with and without the iBHC technique. Several region-of-interest (ROI) measurements were performed in the inferior wall of the left ventricle (LV), an area prone to beam-hardening, as well as other myocardial regions. Coronary contrast-to-noise (CNR) and signal-to-noise ratios (SNR) were calculated. Two radiologists assessed subjective image quality. Results: The iBHC algorithm generally increased myocardial attenuation in all ROIs (P < 0.566); however, the increase was significantly more distinct in beam-hardening prone areas such as the inferior LV (increase, +13.9 HU, +18.6%, P < 0.001), compared to the remaining myocardium (increase, +4.4 HU, +4.5%, P < 0.003). While no significant difference was found for image noise (P < 0.092), greater CNR and SNR values for the left main coronary artery (increase, +20.7% and +17.3%, respectively) were found using the iBHC algorithm (both with P < 0.001). Subjective image quality was comparable between both image series (P = 0.217). Conclusion: The iBHC post-processing algorithm leads to significantly reduced beam-hardening while providing improved objective and equivalent subjective image quality in 70-kV cCTA.

Beam-hardening in 70-kV Coronary CT angiography. Artifact reduction using an advanced post-processing algorithm / Albrecht, M. H.; Bickford, M. W.; Schoepf, U. J.; Tesche, C.; De Santis, D.; Eid, M.; Jacobs, B. E.; Duguay, T. M.; Schmidt, B. T.; Canstein, C.; Varga-Szemes, A.; Leithner, D.; Martin, S.; Vogl, T. J.; De Cecco, C. N.. - In: EUROPEAN JOURNAL OF RADIOLOGY. - ISSN 0720-048X. - 101:(2018), pp. 111-117. [10.1016/j.ejrad.2018.02.016]

Beam-hardening in 70-kV Coronary CT angiography. Artifact reduction using an advanced post-processing algorithm

De Santis D.;
2018

Abstract

Purpose: To investigate the effect of an iterative beam-hardening correction algorithm (iBHC) on artifact reduction and image quality in coronary CT angiography (cCTA) with low tube voltage. Material and methods: Thirty-six patients (17 male, mean age, 57.3 ± 14.5 years) were prospectively enrolled in this IRB-approved study and underwent 70-kV cCTA using a third-generation dual-source CT scanner. Images were reconstructed using a standard algorithm (Bv36) both with and without the iBHC technique. Several region-of-interest (ROI) measurements were performed in the inferior wall of the left ventricle (LV), an area prone to beam-hardening, as well as other myocardial regions. Coronary contrast-to-noise (CNR) and signal-to-noise ratios (SNR) were calculated. Two radiologists assessed subjective image quality. Results: The iBHC algorithm generally increased myocardial attenuation in all ROIs (P < 0.566); however, the increase was significantly more distinct in beam-hardening prone areas such as the inferior LV (increase, +13.9 HU, +18.6%, P < 0.001), compared to the remaining myocardium (increase, +4.4 HU, +4.5%, P < 0.003). While no significant difference was found for image noise (P < 0.092), greater CNR and SNR values for the left main coronary artery (increase, +20.7% and +17.3%, respectively) were found using the iBHC algorithm (both with P < 0.001). Subjective image quality was comparable between both image series (P = 0.217). Conclusion: The iBHC post-processing algorithm leads to significantly reduced beam-hardening while providing improved objective and equivalent subjective image quality in 70-kV cCTA.
2018
artifact reduction; beam-hardening; cardiac ct; infarct evaluation; perfusion assessment
01 Pubblicazione su rivista::01a Articolo in rivista
Beam-hardening in 70-kV Coronary CT angiography. Artifact reduction using an advanced post-processing algorithm / Albrecht, M. H.; Bickford, M. W.; Schoepf, U. J.; Tesche, C.; De Santis, D.; Eid, M.; Jacobs, B. E.; Duguay, T. M.; Schmidt, B. T.; Canstein, C.; Varga-Szemes, A.; Leithner, D.; Martin, S.; Vogl, T. J.; De Cecco, C. N.. - In: EUROPEAN JOURNAL OF RADIOLOGY. - ISSN 0720-048X. - 101:(2018), pp. 111-117. [10.1016/j.ejrad.2018.02.016]
File allegati a questo prodotto
File Dimensione Formato  
Albrecht_Beam-hardening_2018.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 2.66 MB
Formato Adobe PDF
2.66 MB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1634197
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 5
social impact