This paper experimentally validates the capability of a microwave prototype device to localize hemorrhages and ischemias within the brain as well as proposes an innovative calibration technique based on the measured data. In the reported experiments, a 3-D human-like head phantom is considered, where the brain is represented either with a homogeneous liquid mimicking brain dielectric properties or with ex vivo calf brains. The microwave imaging (MWI) system works at 1 GHz, and it is realized with a low-complexity architecture formed by an array of twenty-four printed monopole antennas. Each antenna is embedded into the “brick” of a semi-flexible dielectric matching medium, and it is positioned conformal to the head upper part. The imaging algorithm exploits a differential approach and provides 3-D images of the brain region. It employs the singular value decomposition of the discretized scattering operator obtained via accurate numerical models. The MWI system analysis shows promising reconstruction results and extends the device validation.

Experimental validation of a microwave system for brain stroke 3-d imaging / Rodriguez-Duarte, D. O.; Tobon Vasquez, J. A.; Scapaticci, R.; Turvani, G.; Cavagnaro, M.; Casu, M. R.; Crocco, L.; Vipiana, F.. - In: DIAGNOSTICS. - ISSN 2075-4418. - 11:7(2021), pp. 1-18. [10.3390/diagnostics11071232]

Experimental validation of a microwave system for brain stroke 3-d imaging

Cavagnaro M.;
2021

Abstract

This paper experimentally validates the capability of a microwave prototype device to localize hemorrhages and ischemias within the brain as well as proposes an innovative calibration technique based on the measured data. In the reported experiments, a 3-D human-like head phantom is considered, where the brain is represented either with a homogeneous liquid mimicking brain dielectric properties or with ex vivo calf brains. The microwave imaging (MWI) system works at 1 GHz, and it is realized with a low-complexity architecture formed by an array of twenty-four printed monopole antennas. Each antenna is embedded into the “brick” of a semi-flexible dielectric matching medium, and it is positioned conformal to the head upper part. The imaging algorithm exploits a differential approach and provides 3-D images of the brain region. It employs the singular value decomposition of the discretized scattering operator obtained via accurate numerical models. The MWI system analysis shows promising reconstruction results and extends the device validation.
2021
atenna array; biomedical imaging; brain stroke monitoring; hemorrhagic stroke; ischemic stroke; microwave imaging
01 Pubblicazione su rivista::01a Articolo in rivista
Experimental validation of a microwave system for brain stroke 3-d imaging / Rodriguez-Duarte, D. O.; Tobon Vasquez, J. A.; Scapaticci, R.; Turvani, G.; Cavagnaro, M.; Casu, M. R.; Crocco, L.; Vipiana, F.. - In: DIAGNOSTICS. - ISSN 2075-4418. - 11:7(2021), pp. 1-18. [10.3390/diagnostics11071232]
File allegati a questo prodotto
File Dimensione Formato  
Rodriguezduarte_Experimental_2021.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 629.17 kB
Formato Adobe PDF
629.17 kB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1613455
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 32
  • ???jsp.display-item.citation.isi??? 25
social impact