Background: The traditional technique for subcutaneous implantable cardioverter defibrillator (S-ICD) implantation involves three incisions and a subcutaneous (SC) pocket. An intermuscular (IM) 2-incision technique has been recently adopted. Aims: We assessed acute defibrillation efficacy (DE) of S-ICD (DE ≤65 J) according to the implantation technique. Methods: We analyzed consecutive patients who underwent S-ICD implantation and DE testing at 53 Italian centers. Regression analysis was used to determine the association between DFT and implantation technique. Results: A total of 805 patients were enrolled. Four groups were assessed: IM + 2 incisions (n = 546), SC + 2 incisions (n = 133), SC + 3 incisions (n = 111), and IM + 3 incisions (n = 15). DE was ≤65 J in 782 (97.1%) patients. Patients with DE ≤65 J showed a trend towards lower body mass index (25.1 vs. 26.5; p = .12), were less frequently on antiarrhythmic drugs (13% vs. 26%; p = .06) and more commonly underwent implantation with the 2-incision technique (85% vs. 70%; p = .04). The IM + 2-incision technique showed the lowest defibrillation failure rate (2.2%) and shock impedance (66 Ohm, interquartile range: 57-77). On multivariate analysis, the 2-incision technique was associated with a lower incidence of shock failure (hazard ratio: 0.305; 95% confidence interval: 0.102-0.907; p = .033). Shock impedance was lower with the IM than with the SC approach (66 vs. 70 Ohm p = .002) and with the 2-incision than the 3-incision technique (67 vs. 72 Ohm; p = .006). Conclusions: In a large population of S-ICD patients, we observed a high defibrillation success rate. The IM + 2-incision technique provides lower shock impedance and a higher likelihood of successful defibrillation.

Acute shock efficacy of the subcutaneous implantable cardioverter-defibrillator according to the implantation technique / Francia, Pietro; Adduci, Carmen; Angeletti, Andrea; Ottaviano, Luca; Perrotta, Laura; De Vivo, Stefano; Bongiorni, Maria Grazia; Migliore, Federico; Russo, Antonio Dello; De Filippo, Paolo; Caravati, Fabrizio; Nigro, Gerardo; Palmisano, Pietro; Viani, Stefano; D'Onofrio, Antonio; Lovecchio, Mariolina; Valsecchi, Sergio; Ziacchi, Matteo. - In: JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY. - ISSN 1045-3873. - 32:6(2021), pp. 1695-1703. [10.1111/jce.15081]

Acute shock efficacy of the subcutaneous implantable cardioverter-defibrillator according to the implantation technique

Francia, Pietro
;
Adduci, Carmen;
2021

Abstract

Background: The traditional technique for subcutaneous implantable cardioverter defibrillator (S-ICD) implantation involves three incisions and a subcutaneous (SC) pocket. An intermuscular (IM) 2-incision technique has been recently adopted. Aims: We assessed acute defibrillation efficacy (DE) of S-ICD (DE ≤65 J) according to the implantation technique. Methods: We analyzed consecutive patients who underwent S-ICD implantation and DE testing at 53 Italian centers. Regression analysis was used to determine the association between DFT and implantation technique. Results: A total of 805 patients were enrolled. Four groups were assessed: IM + 2 incisions (n = 546), SC + 2 incisions (n = 133), SC + 3 incisions (n = 111), and IM + 3 incisions (n = 15). DE was ≤65 J in 782 (97.1%) patients. Patients with DE ≤65 J showed a trend towards lower body mass index (25.1 vs. 26.5; p = .12), were less frequently on antiarrhythmic drugs (13% vs. 26%; p = .06) and more commonly underwent implantation with the 2-incision technique (85% vs. 70%; p = .04). The IM + 2-incision technique showed the lowest defibrillation failure rate (2.2%) and shock impedance (66 Ohm, interquartile range: 57-77). On multivariate analysis, the 2-incision technique was associated with a lower incidence of shock failure (hazard ratio: 0.305; 95% confidence interval: 0.102-0.907; p = .033). Shock impedance was lower with the IM than with the SC approach (66 vs. 70 Ohm p = .002) and with the 2-incision than the 3-incision technique (67 vs. 72 Ohm; p = .006). Conclusions: In a large population of S-ICD patients, we observed a high defibrillation success rate. The IM + 2-incision technique provides lower shock impedance and a higher likelihood of successful defibrillation.
2021
icd; conversion; defibrillation test; safety; subcutaneous
01 Pubblicazione su rivista::01a Articolo in rivista
Acute shock efficacy of the subcutaneous implantable cardioverter-defibrillator according to the implantation technique / Francia, Pietro; Adduci, Carmen; Angeletti, Andrea; Ottaviano, Luca; Perrotta, Laura; De Vivo, Stefano; Bongiorni, Maria Grazia; Migliore, Federico; Russo, Antonio Dello; De Filippo, Paolo; Caravati, Fabrizio; Nigro, Gerardo; Palmisano, Pietro; Viani, Stefano; D'Onofrio, Antonio; Lovecchio, Mariolina; Valsecchi, Sergio; Ziacchi, Matteo. - In: JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY. - ISSN 1045-3873. - 32:6(2021), pp. 1695-1703. [10.1111/jce.15081]
File allegati a questo prodotto
File Dimensione Formato  
Francia_Acute-shock-efficacy_2021.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 891.26 kB
Formato Adobe PDF
891.26 kB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1548688
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 5
social impact