The accuracy, reliability and linearity in two hand-portable lactate analyzers, the Accutrend Lactate® (AL) and the Lactate Pro™ (LP) versus the EBIO® plus analyzer (EP) were evaluated. For accuracy, duplicate samples recorded on both the AL and LP revealed an overall average difference versus EP (P < 0.05). The limits of agreement between AL and EP were -0.7 to +1.0, and-1.3 to +1.5 mM between LP and EP. Reliability of AL and LP was assessed at different lactate concentrations; coefficient of variation ranged between 1.8 and 3.3% for AL and between 2.8 and 5.0% for LP. AL and LP had a good reliability for intra-, inter-analyzers, and between test strips (ICC r = 0.999). The linearity was determined versus the EP as reference. The slope coefficient of AL (1.0394) was closer to 1 than that of LP (1.1053). On these bases, AL and LP can be individually considered suitable for the sports research field. © Springer-Verlag 2009.

Accuracy, reliability, linearity of Accutrend and Lactate Pro versus EBIO plus analyzer / Baldari, C.; Bonavolonta, V.; Emerenziani, G. P.; Gallotta, M. C.; Guidetti, L.; Silva, A. J.. - In: EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY. - ISSN 1439-6319. - 107:1(2009), pp. 105-111. [10.1007/s00421-009-1107-5]

Accuracy, reliability, linearity of Accutrend and Lactate Pro versus EBIO plus analyzer

Emerenziani G. P.;Gallotta M. C.;
2009

Abstract

The accuracy, reliability and linearity in two hand-portable lactate analyzers, the Accutrend Lactate® (AL) and the Lactate Pro™ (LP) versus the EBIO® plus analyzer (EP) were evaluated. For accuracy, duplicate samples recorded on both the AL and LP revealed an overall average difference versus EP (P < 0.05). The limits of agreement between AL and EP were -0.7 to +1.0, and-1.3 to +1.5 mM between LP and EP. Reliability of AL and LP was assessed at different lactate concentrations; coefficient of variation ranged between 1.8 and 3.3% for AL and between 2.8 and 5.0% for LP. AL and LP had a good reliability for intra-, inter-analyzers, and between test strips (ICC r = 0.999). The linearity was determined versus the EP as reference. The slope coefficient of AL (1.0394) was closer to 1 than that of LP (1.1053). On these bases, AL and LP can be individually considered suitable for the sports research field. © Springer-Verlag 2009.
2009
Accuracy; Linearity; Portable lactate analyzers; Reliability; Blood Chemical Analysis; Equipment Design; Equipment Failure Analysis; Female; Humans; Lactic Acid; Male; Reproducibility of Results; Sensitivity and Specificity
01 Pubblicazione su rivista::01a Articolo in rivista
Accuracy, reliability, linearity of Accutrend and Lactate Pro versus EBIO plus analyzer / Baldari, C.; Bonavolonta, V.; Emerenziani, G. P.; Gallotta, M. C.; Guidetti, L.; Silva, A. J.. - In: EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY. - ISSN 1439-6319. - 107:1(2009), pp. 105-111. [10.1007/s00421-009-1107-5]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1542693
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 35
  • Scopus 128
  • ???jsp.display-item.citation.isi??? 116
social impact