Purpose This study aimed to examine the validity, reliability and minimum detectable change (MDC) of the Cosmed K5 in breath by breath (BxB) mode, against VacuMed metabolic simulator. Intra and inter-units reliability was also assessed. Methods Fourteen metabolic rates (from 0.9 to 4 L.min-1) were reproduced by a VacuMed system and pulmonary ventilation (VE), oxygen consumption (VO 2 ) and carbon dioxide production (VCO 2 ) were measured by two different K5 units. Validity was assessed by ordinary least products (OLP) regression analysis, Bland-Altman plots, intraclass correlation coefficients (ICC), mean percentage differences, technical errors (TE) and MDC for VE, VO 2 , and VCO 2 . Intra- and inter-K5 reliability was evaluated by absolute percentage differences between measurements (MAPE), ICCs, TE, and MDC. Results Validity analysis from OLP regression data and Bland- Altman plots indicated high agreement between K5 and simulator. ICC values were excellent for all variables (>0.99). Mean percentage differences in VE (-0.50%, p = 0.11), VO 2 (-0.04%, p = 0.80), and VCO 2 (-1.03%, p = 0.09) showed no significant bias. The technical error (TE) ranged from 0.73% to 1.34% (VE and VCO 2 respectively). MDC were lower than 4% (VE = 2.0%, VO 2 = 3.8%, VCO 2 = 3.7%). The intra and inter K5 reliability assessment reveled excellent ICCs (>0.99), MAPE <2% (no significant differences between trials), TE < or around 1%, MDC

Validity, reliability and minimum detectable change of COSMED K5 portable gas exchange system in breath-by-breath mode / Guidetti, L.; Meucci, M.; Bolletta, F.; Emerenziani, G. P.; Gallotta, M. C.; Baldari, C.. - In: PLOS ONE. - ISSN 1932-6203. - 13:12(2018). [10.1371/journal.pone.0209925]

Validity, reliability and minimum detectable change of COSMED K5 portable gas exchange system in breath-by-breath mode

Emerenziani G. P.;Gallotta M. C.;
2018

Abstract

Purpose This study aimed to examine the validity, reliability and minimum detectable change (MDC) of the Cosmed K5 in breath by breath (BxB) mode, against VacuMed metabolic simulator. Intra and inter-units reliability was also assessed. Methods Fourteen metabolic rates (from 0.9 to 4 L.min-1) were reproduced by a VacuMed system and pulmonary ventilation (VE), oxygen consumption (VO 2 ) and carbon dioxide production (VCO 2 ) were measured by two different K5 units. Validity was assessed by ordinary least products (OLP) regression analysis, Bland-Altman plots, intraclass correlation coefficients (ICC), mean percentage differences, technical errors (TE) and MDC for VE, VO 2 , and VCO 2 . Intra- and inter-K5 reliability was evaluated by absolute percentage differences between measurements (MAPE), ICCs, TE, and MDC. Results Validity analysis from OLP regression data and Bland- Altman plots indicated high agreement between K5 and simulator. ICC values were excellent for all variables (>0.99). Mean percentage differences in VE (-0.50%, p = 0.11), VO 2 (-0.04%, p = 0.80), and VCO 2 (-1.03%, p = 0.09) showed no significant bias. The technical error (TE) ranged from 0.73% to 1.34% (VE and VCO 2 respectively). MDC were lower than 4% (VE = 2.0%, VO 2 = 3.8%, VCO 2 = 3.7%). The intra and inter K5 reliability assessment reveled excellent ICCs (>0.99), MAPE <2% (no significant differences between trials), TE < or around 1%, MDC
2018
Breath Tests; Carbon Dioxide; Humans; Oxygen; Oxygen Consumption; Pulmonary Gas Exchange
01 Pubblicazione su rivista::01a Articolo in rivista
Validity, reliability and minimum detectable change of COSMED K5 portable gas exchange system in breath-by-breath mode / Guidetti, L.; Meucci, M.; Bolletta, F.; Emerenziani, G. P.; Gallotta, M. C.; Baldari, C.. - In: PLOS ONE. - ISSN 1932-6203. - 13:12(2018). [10.1371/journal.pone.0209925]
File allegati a questo prodotto
File Dimensione Formato  
Guidetti_Validity_2018.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 937.69 kB
Formato Adobe PDF
937.69 kB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1542420
Citazioni
  • ???jsp.display-item.citation.pmc??? 22
  • Scopus 71
  • ???jsp.display-item.citation.isi??? 69
social impact