Giant Magnetoresistance (GMR) sensors have taken an important role in magnetic field sensing applications thanks to their small size, high sensitivity, large frequency response, low power and relatively low cost. A metrological characterization of a GMR magnetometer has highlighted the contributes to the measurement uncertainty due to hysteresis, non linearity and dependence from temperature. Starting from these knowledge, an automatic procedure for calibration and adjustment of GMR magnetometer has been set up. This procedure, based on a mixed hardware-software solution, allows a deep reduction of the measurement uncertainty for both AC and DC magnetic field strength measurements using GMR

Improving GMR magnetometer sensor uncertainty by implementing an automatic procedure for calibration and adjustment / Bernieri, A.; Ferrigno, L.; Laracca, M.; Tamburrino, A.. - (2007), pp. 1-6. (Intervento presentato al convegno 2007 IEEE Instrumentation and measurement Technology, IMTC 2007 tenutosi a Warsaw; Poland) [10.1109/IMTC.2007.379174].

Improving GMR magnetometer sensor uncertainty by implementing an automatic procedure for calibration and adjustment

LARACCA M.;
2007

Abstract

Giant Magnetoresistance (GMR) sensors have taken an important role in magnetic field sensing applications thanks to their small size, high sensitivity, large frequency response, low power and relatively low cost. A metrological characterization of a GMR magnetometer has highlighted the contributes to the measurement uncertainty due to hysteresis, non linearity and dependence from temperature. Starting from these knowledge, an automatic procedure for calibration and adjustment of GMR magnetometer has been set up. This procedure, based on a mixed hardware-software solution, allows a deep reduction of the measurement uncertainty for both AC and DC magnetic field strength measurements using GMR
2007
2007 IEEE Instrumentation and measurement Technology, IMTC 2007
measurement; GMR; sensor; calibration
04 Pubblicazione in atti di convegno::04b Atto di convegno in volume
Improving GMR magnetometer sensor uncertainty by implementing an automatic procedure for calibration and adjustment / Bernieri, A.; Ferrigno, L.; Laracca, M.; Tamburrino, A.. - (2007), pp. 1-6. (Intervento presentato al convegno 2007 IEEE Instrumentation and measurement Technology, IMTC 2007 tenutosi a Warsaw; Poland) [10.1109/IMTC.2007.379174].
File allegati a questo prodotto
File Dimensione Formato  
Bernieri_Improving GMR_2007.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 3.72 MB
Formato Adobe PDF
3.72 MB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.