OBJECTIVES: To compare device success and paravalvular leak rates of 3 new-generation transcatheter aortic valve replacement devices in patients with bicuspid aortic valve stenosis and to test their biomechanical performance in a computer-based simulation model of aortic root with increasing ellipticity. METHODS: This retrospective multicentre study included 56 bicuspid aortic valve patients undergoing transcatheter aortic valve replacement with new-generation devices: Lotus/Lotus Edge (N = 15; 27%), Evolut-R (N = 20; 36%) and ACURATE neo (N = 21; 37%). Three virtual simulation models of aortic root with increasing index of eccentricity (0-0.25-0.5) were implemented. Stress distribution, stent-root contact area and paravalvular orifice area were computed. RESULTS: Device success was achieved in 43/56 patients (77%) with comparable rates among Lotus (87%), Evolut-R (60%) and ACURATE neo (86%; P = 0.085). Moderate paravalvular leak rate was significantly lower in the Lotus group as compared to Evolut-R group (0% vs 30%; P = 0.027) and comparable to the ACURATE neo group (0% vs 10%; P = 0.33). By index of eccentricity = 0.5, Lotus showed a uniform and symmetric pattern of stress distribution with absent paravalvular orifice area, ACURATE neo showed a mild asymmetry with small paravalvular orifice area (1.1 mm2), whereas a severely asymmetric pattern was evident with Evolut-R, resulting in a large paravalvular orifice area (12.0 mm2). CONCLUSIONS: Transcatheter aortic valve replacement in bicuspid aortic valve patients with new-generation devices showed comparable device success rates. Lotus showed moderate paravalvular leak rate comparable to that of ACURATE neo and significantly lower than Evolut-R. On simulation, Lotus and ACURATE neo showed optimal adaptability to elliptic anatomies as compared to Evolut-R.

Outcome of transcatheter aortic valve replacement in bicuspid aortic valve stenosis with new-generation devices / Gorla, R.; Casenghi, M.; Finotello, A.; De Marco, F.; Morganti, S.; Regazzoli, D.; Bianchi, G.; Acerbi, E.; Popolo Rubbio, A.; Brambilla, N.; Testa, L.; Castriota, F.; Auricchio, F.; Reimers, B.; Bedogni, F.. - In: INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY. - ISSN 1569-9285. - 32:1(2021), pp. 20-28. [10.1093/icvts/ivaa231]

Outcome of transcatheter aortic valve replacement in bicuspid aortic valve stenosis with new-generation devices

Casenghi M.
Secondo
;
Auricchio F.;
2021

Abstract

OBJECTIVES: To compare device success and paravalvular leak rates of 3 new-generation transcatheter aortic valve replacement devices in patients with bicuspid aortic valve stenosis and to test their biomechanical performance in a computer-based simulation model of aortic root with increasing ellipticity. METHODS: This retrospective multicentre study included 56 bicuspid aortic valve patients undergoing transcatheter aortic valve replacement with new-generation devices: Lotus/Lotus Edge (N = 15; 27%), Evolut-R (N = 20; 36%) and ACURATE neo (N = 21; 37%). Three virtual simulation models of aortic root with increasing index of eccentricity (0-0.25-0.5) were implemented. Stress distribution, stent-root contact area and paravalvular orifice area were computed. RESULTS: Device success was achieved in 43/56 patients (77%) with comparable rates among Lotus (87%), Evolut-R (60%) and ACURATE neo (86%; P = 0.085). Moderate paravalvular leak rate was significantly lower in the Lotus group as compared to Evolut-R group (0% vs 30%; P = 0.027) and comparable to the ACURATE neo group (0% vs 10%; P = 0.33). By index of eccentricity = 0.5, Lotus showed a uniform and symmetric pattern of stress distribution with absent paravalvular orifice area, ACURATE neo showed a mild asymmetry with small paravalvular orifice area (1.1 mm2), whereas a severely asymmetric pattern was evident with Evolut-R, resulting in a large paravalvular orifice area (12.0 mm2). CONCLUSIONS: Transcatheter aortic valve replacement in bicuspid aortic valve patients with new-generation devices showed comparable device success rates. Lotus showed moderate paravalvular leak rate comparable to that of ACURATE neo and significantly lower than Evolut-R. On simulation, Lotus and ACURATE neo showed optimal adaptability to elliptic anatomies as compared to Evolut-R.
2021
Bicuspid aortic valve; Computational simulation; New-generation devices; Transcatheter aortic valve replacement
01 Pubblicazione su rivista::01a Articolo in rivista
Outcome of transcatheter aortic valve replacement in bicuspid aortic valve stenosis with new-generation devices / Gorla, R.; Casenghi, M.; Finotello, A.; De Marco, F.; Morganti, S.; Regazzoli, D.; Bianchi, G.; Acerbi, E.; Popolo Rubbio, A.; Brambilla, N.; Testa, L.; Castriota, F.; Auricchio, F.; Reimers, B.; Bedogni, F.. - In: INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY. - ISSN 1569-9285. - 32:1(2021), pp. 20-28. [10.1093/icvts/ivaa231]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1488688
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 7
social impact