Objective: To evaluate the long-term results of using the BacJac interspinous device (Pioneer Surgical Technology Inc.) in a series of patients with degenerative lumbar spine disease. Methods: Forty-one patients undergoing lumbar surgery with implantation of a BacJac device from 2009 to 2012 were enrolled in the present study. Patients were evaluated using the Oswestry Disability Scale (ODI). Results: Although all patients showed a significant improvement of the ODI score immediately after surgery, only 41% of patients showed a satisfactory outcome. We observed worse results in the patients operated on at the L3-L4 level and in whom the device was implanted in a segment different from the one where surgical decompression had been performed. Weight gain in the months after surgery was also a poor outcome-influencing factor. Conclusions: This study confirms what is already suggested in the relevant literature regarding the long-term inefficacy of the so-called dynamic stabilization devices.

Long-term Results for the BacJac Interspinous Device in Lumbar Spine Degenerative Disease / Spallone, A.; Lavorato, L.; Belvisi, D.. - In: JOURNAL OF NEUROLOGICAL SURGERY. PART A, CENTRAL EUROPEAN NEUROSURGERY. - ISSN 2193-6315. - 80:1(2018), pp. 3-7. [10.1055/s-0038-1641180]

Long-term Results for the BacJac Interspinous Device in Lumbar Spine Degenerative Disease

Spallone A.;Lavorato L.;Belvisi D.
2018

Abstract

Objective: To evaluate the long-term results of using the BacJac interspinous device (Pioneer Surgical Technology Inc.) in a series of patients with degenerative lumbar spine disease. Methods: Forty-one patients undergoing lumbar surgery with implantation of a BacJac device from 2009 to 2012 were enrolled in the present study. Patients were evaluated using the Oswestry Disability Scale (ODI). Results: Although all patients showed a significant improvement of the ODI score immediately after surgery, only 41% of patients showed a satisfactory outcome. We observed worse results in the patients operated on at the L3-L4 level and in whom the device was implanted in a segment different from the one where surgical decompression had been performed. Weight gain in the months after surgery was also a poor outcome-influencing factor. Conclusions: This study confirms what is already suggested in the relevant literature regarding the long-term inefficacy of the so-called dynamic stabilization devices.
2018
lumbar instability; lumbar stenosis; lumbar surgery; Adult; Aged; Aged, 80 and over; Decompression, Surgical; Female; Humans; Intervertebral Disc Degeneration; Male; Middle Aged; Retrospective Studies; Time Factors; Treatment Outcome; Internal Fixators; Lumbar Vertebrae
01 Pubblicazione su rivista::01a Articolo in rivista
Long-term Results for the BacJac Interspinous Device in Lumbar Spine Degenerative Disease / Spallone, A.; Lavorato, L.; Belvisi, D.. - In: JOURNAL OF NEUROLOGICAL SURGERY. PART A, CENTRAL EUROPEAN NEUROSURGERY. - ISSN 2193-6315. - 80:1(2018), pp. 3-7. [10.1055/s-0038-1641180]
File allegati a questo prodotto
File Dimensione Formato  
Spallone_Long-term Results_2018.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 177.62 kB
Formato Adobe PDF
177.62 kB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1455986
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 4
social impact