Piezoelectric devices are widely used in oral surgical procedures, including implant site preparation. However, little is known about the influence of working movement on temperature elevation in bone. The aim of this study was to assess the effects of two different working cycles on temperature elevation during piezoelectric implant site preparation. Sixty osteotomies at a depth of 10 mm were performed on bone blocks of bovine ribs using a piezoelectric tip with external irrigation (IM1s, Mectron Medical Technology, Carasco, Italy). A mechanical positioning device was used to guarantee reproducible working and measuring conditions. Two different working cycles, of 4 and 6 s, respectively, were tested, including both longitudinal and rotational movements. Temperature was recorded in real time with a fiber optic thermometer and applied pressure was maintained under 150 g. For each test, the highest recorded temperature (Tmax) and the mean temperature recorded from 30 s before to 30 s after the highest recorded temperature (T±30) were extrapolated. Tests duration was also recorded. Both Tmax and T±30 were significantly higher in the ‘6 s cycles’ group than the ‘4 s cycles’ group (42.44 ± 7.3 ◦C vs. 37.24 ± 4.6 ◦C, p = 0.002; 37.24 ± 4.6 ◦C vs. 33.30 ± 3.3 ◦C, p = 0.003). Test duration was also significantly higher using 6 s cycles compared to 4 s cycles (143.17 ± 29.4 s vs. 119.80 ± 36.4 s, p = 0.002). The results of this study indicate that working cycles of 4 s effectively reduce heat generation and working time during piezoelectric implant site preparation

Piezoelectric implant site preparation: influence of handpiece movements on temperature elevation / Lamazza, Luca; Lollobrigida, Marco; Vozza, Iole; Palmieri, Luigi; Stacchi, Claudio; Lombardi, Teresa; De Biase, Alberto. - In: MATERIALS. - ISSN 1996-1944. - 13:18(2020). [10.3390/ma13184072]

Piezoelectric implant site preparation: influence of handpiece movements on temperature elevation

Lamazza, Luca
Primo
;
Lollobrigida, Marco
Secondo
;
Vozza, Iole;De Biase, Alberto
Ultimo
2020

Abstract

Piezoelectric devices are widely used in oral surgical procedures, including implant site preparation. However, little is known about the influence of working movement on temperature elevation in bone. The aim of this study was to assess the effects of two different working cycles on temperature elevation during piezoelectric implant site preparation. Sixty osteotomies at a depth of 10 mm were performed on bone blocks of bovine ribs using a piezoelectric tip with external irrigation (IM1s, Mectron Medical Technology, Carasco, Italy). A mechanical positioning device was used to guarantee reproducible working and measuring conditions. Two different working cycles, of 4 and 6 s, respectively, were tested, including both longitudinal and rotational movements. Temperature was recorded in real time with a fiber optic thermometer and applied pressure was maintained under 150 g. For each test, the highest recorded temperature (Tmax) and the mean temperature recorded from 30 s before to 30 s after the highest recorded temperature (T±30) were extrapolated. Tests duration was also recorded. Both Tmax and T±30 were significantly higher in the ‘6 s cycles’ group than the ‘4 s cycles’ group (42.44 ± 7.3 ◦C vs. 37.24 ± 4.6 ◦C, p = 0.002; 37.24 ± 4.6 ◦C vs. 33.30 ± 3.3 ◦C, p = 0.003). Test duration was also significantly higher using 6 s cycles compared to 4 s cycles (143.17 ± 29.4 s vs. 119.80 ± 36.4 s, p = 0.002). The results of this study indicate that working cycles of 4 s effectively reduce heat generation and working time during piezoelectric implant site preparation
2020
piezoelectric surgery; implant site preparation; heat generation; temperature rise; working movements
01 Pubblicazione su rivista::01a Articolo in rivista
Piezoelectric implant site preparation: influence of handpiece movements on temperature elevation / Lamazza, Luca; Lollobrigida, Marco; Vozza, Iole; Palmieri, Luigi; Stacchi, Claudio; Lombardi, Teresa; De Biase, Alberto. - In: MATERIALS. - ISSN 1996-1944. - 13:18(2020). [10.3390/ma13184072]
File allegati a questo prodotto
File Dimensione Formato  
Lamazza_Piezoeletric_2020.pdf

accesso aperto

Note: https://www.mdpi.com/1996-1944/13/18/4072
Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 1.19 MB
Formato Adobe PDF
1.19 MB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1438114
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 3
social impact