A proper choice of the scintillation crystal can considerably improve the detector performance in nuclear medicine imaging. In this study two promising crystals, CRY018 and CRY019, are characterized in terms of spectrometric and imaging capabilities. Thanks to the same irradiation conditions, it has been possible to compare their performance for a wide range of energies (30 keV to 662 keV) . They have high energy resolution (6.5% and 7.6% at 662 keV for CRY018 and CRY019, respectively), negligible light yield non-proportionality, good position linearity, and intrinsic spatial resolution less than 2 mm. In particular, the performance deviations could be explained from the different attenuation length (CRY018 depth of interaction is higher than CRY019 one, due to their different density).
Imaging performance dependence on crystal absorption properties: the CRY018 and CRY019 comparison / Pellegrini, R.; Camera, F.; Polito, C.; Falconi, R.; Bettiol, M.; Longo, M.; De Vincentis, G.; Indovina, L.; Pani, R.; Frantellizzi, V.. - In: JOURNAL OF INSTRUMENTATION. - ISSN 1748-0221. - 15:5(2020), pp. 1-11. (Intervento presentato al convegno 15th Topical Seminar on Innovative Particle and Radiation Detectors (IPRD) tenutosi a Siena, ITALY) [10.1088/1748-0221/15/05/C05076].
Imaging performance dependence on crystal absorption properties: the CRY018 and CRY019 comparison
Pellegrini R.;Camera F.;Polito C.;Falconi R.;Bettiol M.;Longo M.;De Vincentis G.;Pani R.;Frantellizzi V.Ultimo
2020
Abstract
A proper choice of the scintillation crystal can considerably improve the detector performance in nuclear medicine imaging. In this study two promising crystals, CRY018 and CRY019, are characterized in terms of spectrometric and imaging capabilities. Thanks to the same irradiation conditions, it has been possible to compare their performance for a wide range of energies (30 keV to 662 keV) . They have high energy resolution (6.5% and 7.6% at 662 keV for CRY018 and CRY019, respectively), negligible light yield non-proportionality, good position linearity, and intrinsic spatial resolution less than 2 mm. In particular, the performance deviations could be explained from the different attenuation length (CRY018 depth of interaction is higher than CRY019 one, due to their different density).File | Dimensione | Formato | |
---|---|---|---|
Pellegrini_Imaging_2020.pdf
solo gestori archivio
Tipologia:
Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza:
Tutti i diritti riservati (All rights reserved)
Dimensione
883.02 kB
Formato
Adobe PDF
|
883.02 kB | Adobe PDF | Contatta l'autore |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.