The use of X-ray fluorescence (XRF) scanning systems has become a common practice in many application sectors. In multistratified and heterogeneous samples, the simple analysis of an XRF spectrum as a response of the entire sample is not reliable, so different spectral analysis techniques have been proposed to detect the presence of surface stratification. One commonly studied case is that of gilding, i.e., the presence of a superimposing gold-leaf layer. The observation of changes in the net peak ratios of a single element or of several elements in an XRF spectrum is a well-developed practice, but is still not used in the case of XRF scanning (macro-X-Ray fluorescence scanning, MA-XRF), a technique that can be described as the extrapolation of XRF spot analysis to a second dimension, scanning a sample surface instead. This practice can yield information on the overlaying layer thickness, if some properties of the sample are known—or estimated—beforehand, e.g., the overlapping layer’s chemical composition and the matrix effect contribution from the bulk material (thick ratio). This work proposes the use of an algorithm to calculate the thickness distribution of a superimposing gold layer accurately and automatically through the differential attenuation method by using MA-XRF datasets in a total noninvasive manner. This approach has the clear advantage over the traditional spot sampling of allowing the generation of a surface heightmap to better visualize and interpret the data, as well as a considerably larger sample space. Monte Carlo simulations were used to verify the influence of the medium used to adhere the gold leaves to the substrate and to generate known spectra to assess the algorithm’s accuracy.

Testing the accuracy of the calculation of gold leaf thickness by MC simulations and MA-XRF scanning / Barcellos Lins, Sergio Augusto; Gigante, Giovanni Ettore; Cesareo, Roberto; Ridolfi, Stefano; Brunetti, Antonio. - In: APPLIED SCIENCES. - ISSN 2076-3417. - 10:10(2020), pp. 1-12. [10.3390/app10103582]

Testing the accuracy of the calculation of gold leaf thickness by MC simulations and MA-XRF scanning

Barcellos Lins, Sergio Augusto
;
Gigante, Giovanni Ettore;Ridolfi, Stefano;
2020

Abstract

The use of X-ray fluorescence (XRF) scanning systems has become a common practice in many application sectors. In multistratified and heterogeneous samples, the simple analysis of an XRF spectrum as a response of the entire sample is not reliable, so different spectral analysis techniques have been proposed to detect the presence of surface stratification. One commonly studied case is that of gilding, i.e., the presence of a superimposing gold-leaf layer. The observation of changes in the net peak ratios of a single element or of several elements in an XRF spectrum is a well-developed practice, but is still not used in the case of XRF scanning (macro-X-Ray fluorescence scanning, MA-XRF), a technique that can be described as the extrapolation of XRF spot analysis to a second dimension, scanning a sample surface instead. This practice can yield information on the overlaying layer thickness, if some properties of the sample are known—or estimated—beforehand, e.g., the overlapping layer’s chemical composition and the matrix effect contribution from the bulk material (thick ratio). This work proposes the use of an algorithm to calculate the thickness distribution of a superimposing gold layer accurately and automatically through the differential attenuation method by using MA-XRF datasets in a total noninvasive manner. This approach has the clear advantage over the traditional spot sampling of allowing the generation of a surface heightmap to better visualize and interpret the data, as well as a considerably larger sample space. Monte Carlo simulations were used to verify the influence of the medium used to adhere the gold leaves to the substrate and to generate known spectra to assess the algorithm’s accuracy.
2020
MA-XRF; Monte Carlo simulations; thickness calculation; differential attenuation
01 Pubblicazione su rivista::01a Articolo in rivista
Testing the accuracy of the calculation of gold leaf thickness by MC simulations and MA-XRF scanning / Barcellos Lins, Sergio Augusto; Gigante, Giovanni Ettore; Cesareo, Roberto; Ridolfi, Stefano; Brunetti, Antonio. - In: APPLIED SCIENCES. - ISSN 2076-3417. - 10:10(2020), pp. 1-12. [10.3390/app10103582]
File allegati a questo prodotto
File Dimensione Formato  
BarcellosLins_Testing_2020.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 2.14 MB
Formato Adobe PDF
2.14 MB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1399286
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 17
  • ???jsp.display-item.citation.isi??? 17
social impact