Objectives This in vivo study evaluated the operative torque and preparation time of ProTaper NEXT (Dentsply Maillefer; Ballaigues, Switzerland) and EdgeFile X7 (EdgeEndo; Albuquerque, New Mexico, United States) rotary systems during root canal preparation of maxillary premolars. Materials and Methods Ten double-rooted maxillary premolars with independent canals were selected. Each canal in each tooth was prepared with one of the rotary systems (n = 10), ProTaper NEXT or EdgeFile X7. The instruments were rotated at 300 rpm with maximum torque set at 2 N.cm using an electric motor (KaVo; Biberach, Germany) that automatically recorded torque values at every 1/10th of a second (ds). Statistical Analysis Operative torque (N.cm) and preparation time (s) of the first shaping instrument (size 17/.04) of both rotary systems were recorded and statistically compared using the Mann-Whiney U test with a significance level set at 5%. Results No instrument exhibited flute deformation or underwent intracanal failure. No differences were found between the instruments regarding the maximum (peak) torque values (p > 0.05). EdgeFile X7 17/.04 required significantly less preparation time (3.75 seconds interquartile range [IQR]: 3.2-9.0) than ProTaper NEXT X1 (15.45 seconds IQR: 8.35-21.1) (p < 0.05). The median operative torque values of ProTaper NEXT X1 (0.26 N.cm; IQR: 0.18-0.49) were significantly higher compared with EdgeFile X7 17/.04 (0.09 N.cm; IQR: 0.05-0.17) (p < 0.05). Conclusions Although no difference was found between the median peak torque values of ProTaper NEXT X1 and EdgeFile X7 17/.04 instruments, the operative torque and instrumentation time results were impacted by their different designs and alloys during clinical preparation of root canals.

In vivo evaluation of operative torque generated by two Nickel-Titanium rotary instruments during root canal preparation / Gambarini, G.; Galli, M.; Seracchiani, M.; Di Nardo, D.; Versiani, M. A.; Piasecki, L.; Testarelli, L.. - In: EUROPEAN JOURNAL OF DENTISTRY. - ISSN 1305-7456. - 13:4(2019), pp. 556-562. [10.1055/s-0039-1698369]

In vivo evaluation of operative torque generated by two Nickel-Titanium rotary instruments during root canal preparation

Gambarini G.
Primo
Formal Analysis
;
Galli M.
Secondo
Membro del Collaboration Group
;
Seracchiani M.
Writing – Original Draft Preparation
;
Di Nardo D.
Conceptualization
;
Testarelli L.
Ultimo
Methodology
2019

Abstract

Objectives This in vivo study evaluated the operative torque and preparation time of ProTaper NEXT (Dentsply Maillefer; Ballaigues, Switzerland) and EdgeFile X7 (EdgeEndo; Albuquerque, New Mexico, United States) rotary systems during root canal preparation of maxillary premolars. Materials and Methods Ten double-rooted maxillary premolars with independent canals were selected. Each canal in each tooth was prepared with one of the rotary systems (n = 10), ProTaper NEXT or EdgeFile X7. The instruments were rotated at 300 rpm with maximum torque set at 2 N.cm using an electric motor (KaVo; Biberach, Germany) that automatically recorded torque values at every 1/10th of a second (ds). Statistical Analysis Operative torque (N.cm) and preparation time (s) of the first shaping instrument (size 17/.04) of both rotary systems were recorded and statistically compared using the Mann-Whiney U test with a significance level set at 5%. Results No instrument exhibited flute deformation or underwent intracanal failure. No differences were found between the instruments regarding the maximum (peak) torque values (p > 0.05). EdgeFile X7 17/.04 required significantly less preparation time (3.75 seconds interquartile range [IQR]: 3.2-9.0) than ProTaper NEXT X1 (15.45 seconds IQR: 8.35-21.1) (p < 0.05). The median operative torque values of ProTaper NEXT X1 (0.26 N.cm; IQR: 0.18-0.49) were significantly higher compared with EdgeFile X7 17/.04 (0.09 N.cm; IQR: 0.05-0.17) (p < 0.05). Conclusions Although no difference was found between the median peak torque values of ProTaper NEXT X1 and EdgeFile X7 17/.04 instruments, the operative torque and instrumentation time results were impacted by their different designs and alloys during clinical preparation of root canals.
2019
EdgeFile; nickel-titanium; ProTaper NEXT; rotary instruments; torque
01 Pubblicazione su rivista::01a Articolo in rivista
In vivo evaluation of operative torque generated by two Nickel-Titanium rotary instruments during root canal preparation / Gambarini, G.; Galli, M.; Seracchiani, M.; Di Nardo, D.; Versiani, M. A.; Piasecki, L.; Testarelli, L.. - In: EUROPEAN JOURNAL OF DENTISTRY. - ISSN 1305-7456. - 13:4(2019), pp. 556-562. [10.1055/s-0039-1698369]
File allegati a questo prodotto
File Dimensione Formato  
Gambarini_In-vivo-evaluation_2019.pdf

accesso aperto

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Creative commons
Dimensione 931.63 kB
Formato Adobe PDF
931.63 kB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1383009
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 25
  • ???jsp.display-item.citation.isi??? ND
social impact