Background and aims: Whether Fibroscan thresholds can be immediately adopted for none, some or all other shear wave elastography techniques has not been tested. The aim of the present study was to test the concordance of the findings obtained from 7 of the most recent ultrasound elastography machines with respect to Fibroscan.Methods: Sixteen hepatitis C virus-related patients with fibrosis >= 2 and having reliable results at Fibroscan were investigated in two intercostal spaces using 7 different elastography machines. Coefficients of both precision (an index of data dispersion) and accuracy (an index of bias correction factors expressing different magnitudes of changes in comparison to the reference) were calculated.Results: Median stiffness values differed among the different machines as did coefficients of both precision (range 0.54-0.72) and accuracy (range 0.28-0.87). When the average of the measurements of two intercostal spaces was considered, coefficients of precision significantly increased with all machines (range 0.72-0.90) whereas of accuracy improved more scatteredly and by a smaller degree (range 0.40-0.99).Conclusions: The present results showed only moderate concordance of the majority of elastography machines with the Fibroscan results, preventing the possibility of the immediate universal adoption of Fibroscan thresholds for defining liver fibrosis staging for all new machines. (C) 2017 Editrice Gastroenterologica Italiana S.r.l. Published by Elsevier Ltd. All rights reserved.

Differences in liver stiffness values obtained with new ultrasound elastography machines and Fibroscan: a comparative study / Piscaglia, F.; Salvatore, V.; Mulazzani, L.; Cantisani, V.; Colecchia, Andrea; Di Donato, R.; Felicani, C.; Ferrarini, A.; Gamal, N.; Grasso, V.; Marasco, G.; Mazzotta, E.; Ravaioli, F.; Ruggieri, G.; Serio, I.; Sitouok Nkamgho, J. F.; Serra, C.; Festi, D.; Schiavone, C.; Bolondi, L.. - In: DIGESTIVE AND LIVER DISEASE. - ISSN 1590-8658. - 49:7(2017), pp. 802-808. [10.1016/j.dld.2017.03.001]

Differences in liver stiffness values obtained with new ultrasound elastography machines and Fibroscan: a comparative study

Cantisani V.;COLECCHIA, ANDREA;Di Donato R.;
2017

Abstract

Background and aims: Whether Fibroscan thresholds can be immediately adopted for none, some or all other shear wave elastography techniques has not been tested. The aim of the present study was to test the concordance of the findings obtained from 7 of the most recent ultrasound elastography machines with respect to Fibroscan.Methods: Sixteen hepatitis C virus-related patients with fibrosis >= 2 and having reliable results at Fibroscan were investigated in two intercostal spaces using 7 different elastography machines. Coefficients of both precision (an index of data dispersion) and accuracy (an index of bias correction factors expressing different magnitudes of changes in comparison to the reference) were calculated.Results: Median stiffness values differed among the different machines as did coefficients of both precision (range 0.54-0.72) and accuracy (range 0.28-0.87). When the average of the measurements of two intercostal spaces was considered, coefficients of precision significantly increased with all machines (range 0.72-0.90) whereas of accuracy improved more scatteredly and by a smaller degree (range 0.40-0.99).Conclusions: The present results showed only moderate concordance of the majority of elastography machines with the Fibroscan results, preventing the possibility of the immediate universal adoption of Fibroscan thresholds for defining liver fibrosis staging for all new machines. (C) 2017 Editrice Gastroenterologica Italiana S.r.l. Published by Elsevier Ltd. All rights reserved.
2017
Cirrhosis; fibroscan; liver fibrosis; shear wave elastography; transient elastography; elasticity imaging techniques; female; hepatitis C; humans; liver; liver cirrhosis; male; sensitivity and specificity; statistics nonparametric
01 Pubblicazione su rivista::01a Articolo in rivista
Differences in liver stiffness values obtained with new ultrasound elastography machines and Fibroscan: a comparative study / Piscaglia, F.; Salvatore, V.; Mulazzani, L.; Cantisani, V.; Colecchia, Andrea; Di Donato, R.; Felicani, C.; Ferrarini, A.; Gamal, N.; Grasso, V.; Marasco, G.; Mazzotta, E.; Ravaioli, F.; Ruggieri, G.; Serio, I.; Sitouok Nkamgho, J. F.; Serra, C.; Festi, D.; Schiavone, C.; Bolondi, L.. - In: DIGESTIVE AND LIVER DISEASE. - ISSN 1590-8658. - 49:7(2017), pp. 802-808. [10.1016/j.dld.2017.03.001]
File allegati a questo prodotto
File Dimensione Formato  
Cantisano_Liver-stiffness-values.pdf

accesso aperto

Tipologia: Documento in Post-print (versione successiva alla peer review e accettata per la pubblicazione)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 1.31 MB
Formato Adobe PDF
1.31 MB Adobe PDF

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1319004
Citazioni
  • ???jsp.display-item.citation.pmc??? 11
  • Scopus 53
  • ???jsp.display-item.citation.isi??? 44
social impact