This study evaluated the effect of new motions of the motor TriAuto ZX2 on the cyclic fatigue of endodontic instruments. Vortex Blue 35.06 instruments were divided into four groups (n = 10) and tested for fatigue in a curved artificial canal (90° and 2 mm radius) using the following motions: continuous rotation (CR), Optimum Torque Reverse (OTR) set at 180° and the Optimum Glide Path (OGP), which was tested at 90° and 240°. The time to fracture (TTF) and the lengths of the fractured fragments were recorded. The mean TTF was significantly different among the groups (anova, P < 0.05): OGP 90° (213.39 ± 27.45), OTR 180° (121.24 ± 17.03), OGP 240° (45.24 ± 5.61) and CR (8.43 ± 1.27). Weibull analysis confirmed the shortest life expectancy for CR and the longest survival for OGP at 90°. The resistance to fatigue was affected by motions and pre-set angles. The proprietary movements that are currently available for endodontic instruments were classified according to their kinematics.
Classification and cyclic fatigue evaluation of new kinematics for endodontic instruments / Gambarini, G.; Piasecki, L.; Miccoli, G.; Gaimari, G.; Di Giorgio, R.; Di Nardo, D.; Azim, A. A.; Testarelli, L.. - In: AUSTRALIAN ENDODONTIC JOURNAL. - ISSN 1329-1947. - 45:2(2019), pp. 154-162. [10.1111/aej.12294]
Classification and cyclic fatigue evaluation of new kinematics for endodontic instruments
Gambarini G.Primo
Validation
;Miccoli G.Data Curation
;Gaimari G.;Di Giorgio R.;Di Nardo D.Methodology
;Testarelli L.Ultimo
Conceptualization
2019
Abstract
This study evaluated the effect of new motions of the motor TriAuto ZX2 on the cyclic fatigue of endodontic instruments. Vortex Blue 35.06 instruments were divided into four groups (n = 10) and tested for fatigue in a curved artificial canal (90° and 2 mm radius) using the following motions: continuous rotation (CR), Optimum Torque Reverse (OTR) set at 180° and the Optimum Glide Path (OGP), which was tested at 90° and 240°. The time to fracture (TTF) and the lengths of the fractured fragments were recorded. The mean TTF was significantly different among the groups (anova, P < 0.05): OGP 90° (213.39 ± 27.45), OTR 180° (121.24 ± 17.03), OGP 240° (45.24 ± 5.61) and CR (8.43 ± 1.27). Weibull analysis confirmed the shortest life expectancy for CR and the longest survival for OGP at 90°. The resistance to fatigue was affected by motions and pre-set angles. The proprietary movements that are currently available for endodontic instruments were classified according to their kinematics.File | Dimensione | Formato | |
---|---|---|---|
Gambarini_Classification_2019.pdf
solo gestori archivio
Tipologia:
Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza:
Tutti i diritti riservati (All rights reserved)
Dimensione
1.57 MB
Formato
Adobe PDF
|
1.57 MB | Adobe PDF | Contatta l'autore |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.