Endovascular aneurysm sealing (EVAS) using the Nellix™ System was introduced in clinical practice with the aim of reducing the incidence of complications such as migration, endoleaks, and reinterventions after conventional endovascular aneurysm repair (EVAR). Although, initial efficacy data on this device have been encouraging, EVAS has also demonstrated to undergo adverse events. Herein, we report a case of Nellix graft explant due to endobags shrinkage after air bubble reabsorption leading to proximal type I A endoleak and stent migration. The focus of this article is on the importance of a more assiduous surveillance of this new device, in particular in those cases with air into the endobags immediately after the procedure; this surveillance should be aimed to timely identify complications which can otherwise lead to consequences that require open conversion.

Nellix endovascular aortic sealing endoprosthesis late explantation for concomitant type I endoleak and stent frames proximal caudal migration / Martinelli, O.; Fresilli, M.; Irace, L.; Gossetti, B.; Di Girolamo, A.. - In: ANNALS OF VASCULAR SURGERY. - ISSN 0890-5096. - 19:(2019). [10.1016/j.avsg.2019.03.023]

Nellix endovascular aortic sealing endoprosthesis late explantation for concomitant type I endoleak and stent frames proximal caudal migration

Martinelli O.;Irace L.;Gossetti B.;Di Girolamo A.
2019

Abstract

Endovascular aneurysm sealing (EVAS) using the Nellix™ System was introduced in clinical practice with the aim of reducing the incidence of complications such as migration, endoleaks, and reinterventions after conventional endovascular aneurysm repair (EVAR). Although, initial efficacy data on this device have been encouraging, EVAS has also demonstrated to undergo adverse events. Herein, we report a case of Nellix graft explant due to endobags shrinkage after air bubble reabsorption leading to proximal type I A endoleak and stent migration. The focus of this article is on the importance of a more assiduous surveillance of this new device, in particular in those cases with air into the endobags immediately after the procedure; this surveillance should be aimed to timely identify complications which can otherwise lead to consequences that require open conversion.
2019
Endovascular aneurysm sealing; Nellix™ System; type I A endoleak; stent migration; open conversion
01 Pubblicazione su rivista::01a Articolo in rivista
Nellix endovascular aortic sealing endoprosthesis late explantation for concomitant type I endoleak and stent frames proximal caudal migration / Martinelli, O.; Fresilli, M.; Irace, L.; Gossetti, B.; Di Girolamo, A.. - In: ANNALS OF VASCULAR SURGERY. - ISSN 0890-5096. - 19:(2019). [10.1016/j.avsg.2019.03.023]
File allegati a questo prodotto
File Dimensione Formato  
Martinelli_postprint_Nellix_2019.pdf

Open Access dal 24/03/2020

Tipologia: Documento in Post-print (versione successiva alla peer review e accettata per la pubblicazione)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 852.05 kB
Formato Adobe PDF
852.05 kB Adobe PDF
Martinelli_Nellix_2019.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 3.35 MB
Formato Adobe PDF
3.35 MB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1307865
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 6
social impact