Endovascular aneurysm sealing (EVAS) using the Nellix™ System was introduced in clinical practice with the aim of reducing the incidence of complications such as migration, endoleaks, and reinterventions after conventional endovascular aneurysm repair (EVAR). Although, initial efficacy data on this device have been encouraging, EVAS has also demonstrated to undergo adverse events. Herein, we report a case of Nellix graft explant due to endobags shrinkage after air bubble reabsorption leading to proximal type I A endoleak and stent migration. The focus of this article is on the importance of a more assiduous surveillance of this new device, in particular in those cases with air into the endobags immediately after the procedure; this surveillance should be aimed to timely identify complications which can otherwise lead to consequences that require open conversion.
Nellix endovascular aortic sealing endoprosthesis late explantation for concomitant type I endoleak and stent frames proximal caudal migration / Martinelli, O.; Fresilli, M.; Irace, L.; Gossetti, B.; Di Girolamo, A.. - In: ANNALS OF VASCULAR SURGERY. - ISSN 0890-5096. - 19:(2019). [10.1016/j.avsg.2019.03.023]
Nellix endovascular aortic sealing endoprosthesis late explantation for concomitant type I endoleak and stent frames proximal caudal migration
Martinelli O.;Irace L.;Gossetti B.;Di Girolamo A.
2019
Abstract
Endovascular aneurysm sealing (EVAS) using the Nellix™ System was introduced in clinical practice with the aim of reducing the incidence of complications such as migration, endoleaks, and reinterventions after conventional endovascular aneurysm repair (EVAR). Although, initial efficacy data on this device have been encouraging, EVAS has also demonstrated to undergo adverse events. Herein, we report a case of Nellix graft explant due to endobags shrinkage after air bubble reabsorption leading to proximal type I A endoleak and stent migration. The focus of this article is on the importance of a more assiduous surveillance of this new device, in particular in those cases with air into the endobags immediately after the procedure; this surveillance should be aimed to timely identify complications which can otherwise lead to consequences that require open conversion.File | Dimensione | Formato | |
---|---|---|---|
Martinelli_postprint_Nellix_2019.pdf
Open Access dal 24/03/2020
Tipologia:
Documento in Post-print (versione successiva alla peer review e accettata per la pubblicazione)
Licenza:
Tutti i diritti riservati (All rights reserved)
Dimensione
852.05 kB
Formato
Adobe PDF
|
852.05 kB | Adobe PDF | |
Martinelli_Nellix_2019.pdf
solo gestori archivio
Tipologia:
Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza:
Tutti i diritti riservati (All rights reserved)
Dimensione
3.35 MB
Formato
Adobe PDF
|
3.35 MB | Adobe PDF | Contatta l'autore |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.