This is the first report of a significant time-to-therapy prolongation due to entrapped air. This case emphasizes the importance of ensuring that all the air in the tunneling channel is eliminated once the lead is in place. Also, it represents an insightful example of how the S-ICD automatic sensitivity algorithm works.
Ventricular fibrillation undersensing due to air entrapment in a patient implanted with a subcutaneous cardioverter defibrillator / Adduci, Carmen; Spadoni, Loris; Palano, Francesca; Francia, Pietro. - In: JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY. - ISSN 1045-3873. - Jun 4(2019), pp. 1-2. [10.1111/jce.14005]
Ventricular fibrillation undersensing due to air entrapment in a patient implanted with a subcutaneous cardioverter defibrillator
Adduci, Carmen;Palano, Francesca;Francia, Pietro
2019
Abstract
This is the first report of a significant time-to-therapy prolongation due to entrapped air. This case emphasizes the importance of ensuring that all the air in the tunneling channel is eliminated once the lead is in place. Also, it represents an insightful example of how the S-ICD automatic sensitivity algorithm works.File | Dimensione | Formato | |
---|---|---|---|
Adduci_Ventricular-fibrillation_2019.pdf
solo gestori archivio
Tipologia:
Documento in Post-print (versione successiva alla peer review e accettata per la pubblicazione)
Licenza:
Tutti i diritti riservati (All rights reserved)
Dimensione
469.22 kB
Formato
Adobe PDF
|
469.22 kB | Adobe PDF | Contatta l'autore |
Adduci_Ventricular-fibrillation_2019.pdf
solo gestori archivio
Tipologia:
Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza:
Tutti i diritti riservati (All rights reserved)
Dimensione
680.48 kB
Formato
Adobe PDF
|
680.48 kB | Adobe PDF | Contatta l'autore |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.