Background: In hypertrophic cardiomyopathy (HCM) patients implanted with an implantable cardioverter defibrillator (ICD), clinical outcomes of antitachycardia pacing (ATP) have been poorly explored. In a retrospective analysis of a cohort of consecutive HCM patients implanted with an ICD, we aimed to assess the efficacy, safety, and clinical value of ATP. Methods: The cohort of HCM patients implanted with a transvenous ICD and followed in our center was assessed for device intervention from implantation to last clinical follow-up. Results: Overall 77 patients (45 males; mean age: 46 ± 16 years) were analyzed. After 67 ± 41 months, 24 patients had 49 ventricular tachycardia/fibrillation (VT/VF) appropriately treated (5.8% per year). Among 39 monomorphic VTs, ATP was effective in 27 (success rate: 69%). Mean time from VT onset to ATP delivery was 9.1 ± 4.9 s. The only clinical variable improving ATP success was use of beta-blockers (81% vs 50%; P =.04). Out of 12 ineffectively treated VTs, one was immediately shocked, four self-terminated after 18 ± 12 s, and seven (18%) were accelerated to a new VT. ATP was also delivered for 27 of 42 inappropriately detected episodes and induced two de novo VTs (7%). In the per patient analysis, 14 out 77 (18%) patients had one or more appropriate and effective ATP (3.4% per year), and only six (8%; 1.4% per year) received more than one ATP. Conclusion: ATP is moderately effective for the treatment of monomorphic VTs in HCM patients. However, the rate of appropriate ATP therapies is low, ATP is often prematurely delivered, and ATP-induced arrhythmia degeneration is of concern.

Safety and efficacy of anti-tachycardia pacing in patients with hypertrophic cardiomyopathy implanted with an ICD / Adduci, C.; Semprini, L.; Palano, F.; Musumeci, M. B.; Volpe, M.; Autore, C.; Francia, P.. - In: PACING AND CLINICAL ELECTROPHYSIOLOGY. - ISSN 0147-8389. - 42:6(2019), pp. 610-616. [10.1111/pace.13665]

Safety and efficacy of anti-tachycardia pacing in patients with hypertrophic cardiomyopathy implanted with an ICD

Adduci C.;Semprini L.;Palano F.;Musumeci M. B.;Volpe M.;Autore C.;Francia P.
2019

Abstract

Background: In hypertrophic cardiomyopathy (HCM) patients implanted with an implantable cardioverter defibrillator (ICD), clinical outcomes of antitachycardia pacing (ATP) have been poorly explored. In a retrospective analysis of a cohort of consecutive HCM patients implanted with an ICD, we aimed to assess the efficacy, safety, and clinical value of ATP. Methods: The cohort of HCM patients implanted with a transvenous ICD and followed in our center was assessed for device intervention from implantation to last clinical follow-up. Results: Overall 77 patients (45 males; mean age: 46 ± 16 years) were analyzed. After 67 ± 41 months, 24 patients had 49 ventricular tachycardia/fibrillation (VT/VF) appropriately treated (5.8% per year). Among 39 monomorphic VTs, ATP was effective in 27 (success rate: 69%). Mean time from VT onset to ATP delivery was 9.1 ± 4.9 s. The only clinical variable improving ATP success was use of beta-blockers (81% vs 50%; P =.04). Out of 12 ineffectively treated VTs, one was immediately shocked, four self-terminated after 18 ± 12 s, and seven (18%) were accelerated to a new VT. ATP was also delivered for 27 of 42 inappropriately detected episodes and induced two de novo VTs (7%). In the per patient analysis, 14 out 77 (18%) patients had one or more appropriate and effective ATP (3.4% per year), and only six (8%; 1.4% per year) received more than one ATP. Conclusion: ATP is moderately effective for the treatment of monomorphic VTs in HCM patients. However, the rate of appropriate ATP therapies is low, ATP is often prematurely delivered, and ATP-induced arrhythmia degeneration is of concern.
2019
antitachycardia pacing; hypertrophic cardiomyopathy; implantable cardioverter defibrillator; ventricular arrhythmia
01 Pubblicazione su rivista::01a Articolo in rivista
Safety and efficacy of anti-tachycardia pacing in patients with hypertrophic cardiomyopathy implanted with an ICD / Adduci, C.; Semprini, L.; Palano, F.; Musumeci, M. B.; Volpe, M.; Autore, C.; Francia, P.. - In: PACING AND CLINICAL ELECTROPHYSIOLOGY. - ISSN 0147-8389. - 42:6(2019), pp. 610-616. [10.1111/pace.13665]
File allegati a questo prodotto
File Dimensione Formato  
Adduci_Safety_2019.pdf

solo gestori archivio

Tipologia: Documento in Post-print (versione successiva alla peer review e accettata per la pubblicazione)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 1.78 MB
Formato Adobe PDF
1.78 MB Adobe PDF   Contatta l'autore
Adduci_Safety_2019.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 992.48 kB
Formato Adobe PDF
992.48 kB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1291755
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 10
  • ???jsp.display-item.citation.isi??? 10
social impact