OBJECTIVE: To compare the effect of contrast medium iodine concentration on contrast enhancement, heart rate, and injection pressure when injected at a constant iodine delivery rate in coronary CT angiography (CTA). METHODS: One thousand twenty-four patients scheduled for coronary CTA were prospectively randomized to receive one of four contrast media: iopromide 300 mg I/ml, iohexol 350 mg I/ml, iopromide 370 mg I/ml, or iomeprol 400 mg I/ml. Contrast media were delivered at an equivalent iodine delivery rate of 2.0 g I/s. Intracoronary attenuation was measured and compared (per vessel and per segment). Heart rate before and after contrast media injection was documented. Injection pressure was recorded (n = 403) during contrast medium injection and compared between groups. RESULTS: Intracoronary attenuation values were similar for the different contrast groups. The mean attenuation over all segments ranged between 384 HU for 350 mg I/ml and 395 HU for 400 mg I/ml (p = 0.079). Dose-length product (p = 0.8424), signal-to-noise ratio (all p > 0.05), time to peak (p = 0.324), and changes in heart rate (p = 0.974) were comparable between groups. The peak pressures differed: 197.4 psi for 300 mg I/ml (viscosity 4.6 mPa s), 229.8 psi for 350 mg I/ml (10.4 mPa s), 216.1 psi for 370 mg I/ml (9.5 mPa s), and 243.7 psi for 400 mg I/ml (12.6 mPa s) (p < 0.0001). CONCLUSION: Intravascular attenuation and changes in heart rate are independent of iodine concentration when contrast media are injected at the same iodine delivery rate. Differences in injection pressures are associated with the viscosity of the contrast media. KEY POINTS: • The contrast enhancement in coronary CT angiography is independent of the iodine concentration when contrast media are injected at body temperature (37 °C) with the same iodine delivery rate. • Iodine concentration does not influence the change in heart rate when contrast media are injected at identical iodine delivery rates. • For a fixed iodine delivery rate and contrast temperature, the viscosity of the contrast medium affects the injection pressure.

Impact of iodine concentration and iodine delivery rate on contrast enhancement in coronary CT angiography: a randomized multicenter trial (CT-CON) / Rengo, Marco; Dharampal, Anoeshka; Lubbers, Marisa; Kock, Marc; Wildberger, Joachim E; Das, Marco; Niezen, André; van Tilborg, Fiek; Kofflard, Marcel; Laghi, Andrea; Krestin, Gabriel; Nieman, Koen. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - (2019). [10.1007/s00330-019-06196-7]

Impact of iodine concentration and iodine delivery rate on contrast enhancement in coronary CT angiography: a randomized multicenter trial (CT-CON)

Rengo, Marco
;
Laghi, Andrea;
2019

Abstract

OBJECTIVE: To compare the effect of contrast medium iodine concentration on contrast enhancement, heart rate, and injection pressure when injected at a constant iodine delivery rate in coronary CT angiography (CTA). METHODS: One thousand twenty-four patients scheduled for coronary CTA were prospectively randomized to receive one of four contrast media: iopromide 300 mg I/ml, iohexol 350 mg I/ml, iopromide 370 mg I/ml, or iomeprol 400 mg I/ml. Contrast media were delivered at an equivalent iodine delivery rate of 2.0 g I/s. Intracoronary attenuation was measured and compared (per vessel and per segment). Heart rate before and after contrast media injection was documented. Injection pressure was recorded (n = 403) during contrast medium injection and compared between groups. RESULTS: Intracoronary attenuation values were similar for the different contrast groups. The mean attenuation over all segments ranged between 384 HU for 350 mg I/ml and 395 HU for 400 mg I/ml (p = 0.079). Dose-length product (p = 0.8424), signal-to-noise ratio (all p > 0.05), time to peak (p = 0.324), and changes in heart rate (p = 0.974) were comparable between groups. The peak pressures differed: 197.4 psi for 300 mg I/ml (viscosity 4.6 mPa s), 229.8 psi for 350 mg I/ml (10.4 mPa s), 216.1 psi for 370 mg I/ml (9.5 mPa s), and 243.7 psi for 400 mg I/ml (12.6 mPa s) (p < 0.0001). CONCLUSION: Intravascular attenuation and changes in heart rate are independent of iodine concentration when contrast media are injected at the same iodine delivery rate. Differences in injection pressures are associated with the viscosity of the contrast media. KEY POINTS: • The contrast enhancement in coronary CT angiography is independent of the iodine concentration when contrast media are injected at body temperature (37 °C) with the same iodine delivery rate. • Iodine concentration does not influence the change in heart rate when contrast media are injected at identical iodine delivery rates. • For a fixed iodine delivery rate and contrast temperature, the viscosity of the contrast medium affects the injection pressure.
2019
Computed tomography angiography; contrast media; coronary disease; randomized controlled trial
01 Pubblicazione su rivista::01a Articolo in rivista
Impact of iodine concentration and iodine delivery rate on contrast enhancement in coronary CT angiography: a randomized multicenter trial (CT-CON) / Rengo, Marco; Dharampal, Anoeshka; Lubbers, Marisa; Kock, Marc; Wildberger, Joachim E; Das, Marco; Niezen, André; van Tilborg, Fiek; Kofflard, Marcel; Laghi, Andrea; Krestin, Gabriel; Nieman, Koen. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - (2019). [10.1007/s00330-019-06196-7]
File allegati a questo prodotto
File Dimensione Formato  
Rengo_Impact_2019.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 539.76 kB
Formato Adobe PDF
539.76 kB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1262122
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 10
  • ???jsp.display-item.citation.isi??? 9
social impact