We report an unusual complication of Chocolate nitinol-constraining structure after right superficial femoral artery (SFA) angioplasty. The procedure was performed by vascular surgeons in an operating theater equipped by a portable fluoroscopy unit. Under local anesthesia, by a contralateral approach, a 7F introducer sheath was advanced through the proximal portion of the right common iliac artery. Owing to the severe aorto-iliac vessels calcification, it was not possible to place the introducer sheath into a more distal vessel, as planned. After external iliac artery (EIA) stenting (7 × 80 mm Eluvia), SFA obstruction was intraluminal crossed, and a 6 × 120 mm nitinol-constrained balloon (Chocolate; Medtronic) was advanced in place and inflated. Once the balloon came out, the nitinol-constraining structure was not attached to the balloon surface. Under fluoroscopy, the crashed nitinol mesh was identified at distal edge of previously positioned EIA stent. To prevent mesh migration, it was fixed by covering with a 7 × 40 mm stent. The procedure was then successfully completed, as planned. One-month, postoperative computed tomography angiography showed complete expansion of the stents and no significant residual stenosis (>30%) in EIA, and SFA. Chocolate's mesh was still evident between the stent and the iliac artery wall, in absence of further complications. A 3 months follow-up, patient was still completely asymptomatic for claudication.

Unexpected Complication of a Nitinol-Constrained Balloon Angioplasty (Chocolate) in Rutherford Class 3 Patient Presenting Challenging Aorto-Iliac Anatomy / Mansour, Wassim; Baldassarre, Virgilio; Sirignano, Pasqualino; Capoccia, Laura; Cuozzo, Simone; Speziale, Francesco. - In: ANNALS OF VASCULAR SURGERY. - ISSN 0890-5096. - 53:(2018), pp. 271-271.e5. [10.1016/j.avsg.2018.05.070]

Unexpected Complication of a Nitinol-Constrained Balloon Angioplasty (Chocolate) in Rutherford Class 3 Patient Presenting Challenging Aorto-Iliac Anatomy

Mansour, Wassim;BALDASSARRE, VIRGILIO;Sirignano, Pasqualino;Capoccia, Laura;Speziale, Francesco
2018

Abstract

We report an unusual complication of Chocolate nitinol-constraining structure after right superficial femoral artery (SFA) angioplasty. The procedure was performed by vascular surgeons in an operating theater equipped by a portable fluoroscopy unit. Under local anesthesia, by a contralateral approach, a 7F introducer sheath was advanced through the proximal portion of the right common iliac artery. Owing to the severe aorto-iliac vessels calcification, it was not possible to place the introducer sheath into a more distal vessel, as planned. After external iliac artery (EIA) stenting (7 × 80 mm Eluvia), SFA obstruction was intraluminal crossed, and a 6 × 120 mm nitinol-constrained balloon (Chocolate; Medtronic) was advanced in place and inflated. Once the balloon came out, the nitinol-constraining structure was not attached to the balloon surface. Under fluoroscopy, the crashed nitinol mesh was identified at distal edge of previously positioned EIA stent. To prevent mesh migration, it was fixed by covering with a 7 × 40 mm stent. The procedure was then successfully completed, as planned. One-month, postoperative computed tomography angiography showed complete expansion of the stents and no significant residual stenosis (>30%) in EIA, and SFA. Chocolate's mesh was still evident between the stent and the iliac artery wall, in absence of further complications. A 3 months follow-up, patient was still completely asymptomatic for claudication.
2018
Aged; Angioplasty, Balloon; Computed Tomography Angiography; Femoral Artery; Humans; Iliac Artery; Male; Peripheral Arterial Disease; Prosthesis Design; Prosthesis Failure; Treatment Outcome; Vascular Patency; Alloys; Stents; Surgery; Cardiology and Cardiovascular Medicine
01 Pubblicazione su rivista::01a Articolo in rivista
Unexpected Complication of a Nitinol-Constrained Balloon Angioplasty (Chocolate) in Rutherford Class 3 Patient Presenting Challenging Aorto-Iliac Anatomy / Mansour, Wassim; Baldassarre, Virgilio; Sirignano, Pasqualino; Capoccia, Laura; Cuozzo, Simone; Speziale, Francesco. - In: ANNALS OF VASCULAR SURGERY. - ISSN 0890-5096. - 53:(2018), pp. 271-271.e5. [10.1016/j.avsg.2018.05.070]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1259494
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact