An open resonator compatible with X-band commercial spectrometers for electron paramagnetic resonance was designed and built. In the design process, resonators with different shapes and dimensions were compared by using ad hoc defined cavity figures of merit. Three cylindrical cavities operating on TEX11 modes with X = 1 , 2, 3, were designed taking into account a coupling iris and a slit for microwave magnetic field leakage. The cavities were further improved toward the homogeneity of the 100 kHz modulation magnetic field. A cavity was then built, operating on TE111 mode; the measured electromagnetic parameters were in a good agreement with simulations. Finally, the cavity was inserted in a commercial X-band spectrometer, and measurements have been performed on three paramagnetic samples: 2, 2-diphenyl-1-picrylhydrazyl, and irradiated alanine and bone.

Design and realization of an open EPR resonator at x-band frequencies / Pisa, Stefano; Della Monaca, Sara; Di Trocchio, Luigi; Ficcadenti, Luca; Mostacci, Andrea; Pittella, Erika; Sacco, Giulia; Fattibene, Paola. - In: IEEE TRANSACTIONS ON MAGNETICS. - ISSN 0018-9464. - 55:2(2019), pp. 1-10. [10.1109/TMAG.2018.2886148]

Design and realization of an open EPR resonator at x-band frequencies

Pisa, Stefano
;
Ficcadenti, Luca;Mostacci, Andrea;Pittella, Erika;Sacco, Giulia;
2019

Abstract

An open resonator compatible with X-band commercial spectrometers for electron paramagnetic resonance was designed and built. In the design process, resonators with different shapes and dimensions were compared by using ad hoc defined cavity figures of merit. Three cylindrical cavities operating on TEX11 modes with X = 1 , 2, 3, were designed taking into account a coupling iris and a slit for microwave magnetic field leakage. The cavities were further improved toward the homogeneity of the 100 kHz modulation magnetic field. A cavity was then built, operating on TE111 mode; the measured electromagnetic parameters were in a good agreement with simulations. Finally, the cavity was inserted in a commercial X-band spectrometer, and measurements have been performed on three paramagnetic samples: 2, 2-diphenyl-1-picrylhydrazyl, and irradiated alanine and bone.
2019
Cavity resonators; electron paramagnetic resonance (EPR); Helmholtz coils; Electronic, Optical and Magnetic Materials; Electrical and Electronic Engineering
01 Pubblicazione su rivista::01a Articolo in rivista
Design and realization of an open EPR resonator at x-band frequencies / Pisa, Stefano; Della Monaca, Sara; Di Trocchio, Luigi; Ficcadenti, Luca; Mostacci, Andrea; Pittella, Erika; Sacco, Giulia; Fattibene, Paola. - In: IEEE TRANSACTIONS ON MAGNETICS. - ISSN 0018-9464. - 55:2(2019), pp. 1-10. [10.1109/TMAG.2018.2886148]
File allegati a questo prodotto
File Dimensione Formato  
Pisa_Design-and-realization_2019.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 2.45 MB
Formato Adobe PDF
2.45 MB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1230746
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 3
social impact