Purpose. The purpose of this study was to evaluate the cyclic fatigue of ProTaper nickel-titanium (NiTi) rotary instruments (Tulsa Dental, Tulsa, Okla) after multiple clinical uses. Study design. Two hundred twenty-five ProTaper instruments were divided into 3 groups: A = 75 used as controls, B = 75 used in 2 molars (6-8 canals), and C = 75 used in 4 molars (12-16 canals). The number of rotations to breakage and the fractured tip length were recorded for each file and the mean was calculated by ANOVA. Results. No S1-2 or F1-2-3 instrument separated during intracanal use, even if they were reused for a number of cases (X 4 molar cases). Conclusion. It is clear that prolonged reuse of NiTi rotary instruments strongly affects instruments' fatigue, but our data suggest the hypothesis that other factors (primarily errors and misuse) may be more accountable for intracanal instrument separation. Further studies could assess the cyclic fatigue of each instrument at different levels of the shaft by altering the radius of curvature.

Cyclic fatigue testing of ProTaper NiTi rotary instruments after clinical use / D., Fife; Gambarini, Gianluca; L. R., Britto. - In: ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS. - ISSN 1079-2104. - 97:2(2004), pp. 251-256. [10.1016/j.tripleo.2003.08.010]

Cyclic fatigue testing of ProTaper NiTi rotary instruments after clinical use

GAMBARINI, Gianluca;
2004

Abstract

Purpose. The purpose of this study was to evaluate the cyclic fatigue of ProTaper nickel-titanium (NiTi) rotary instruments (Tulsa Dental, Tulsa, Okla) after multiple clinical uses. Study design. Two hundred twenty-five ProTaper instruments were divided into 3 groups: A = 75 used as controls, B = 75 used in 2 molars (6-8 canals), and C = 75 used in 4 molars (12-16 canals). The number of rotations to breakage and the fractured tip length were recorded for each file and the mean was calculated by ANOVA. Results. No S1-2 or F1-2-3 instrument separated during intracanal use, even if they were reused for a number of cases (X 4 molar cases). Conclusion. It is clear that prolonged reuse of NiTi rotary instruments strongly affects instruments' fatigue, but our data suggest the hypothesis that other factors (primarily errors and misuse) may be more accountable for intracanal instrument separation. Further studies could assess the cyclic fatigue of each instrument at different levels of the shaft by altering the radius of curvature.
2004
01 Pubblicazione su rivista::01a Articolo in rivista
Cyclic fatigue testing of ProTaper NiTi rotary instruments after clinical use / D., Fife; Gambarini, Gianluca; L. R., Britto. - In: ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS. - ISSN 1079-2104. - 97:2(2004), pp. 251-256. [10.1016/j.tripleo.2003.08.010]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/122961
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 8
  • Scopus 93
  • ???jsp.display-item.citation.isi??? 87
social impact