This study investigated the cyclic fatigue resistance of used nickel-titanium rotary instruments, clinically operated by a traditional high-torque motor and a new low-torque electric motor. Fifty instruments (ProFile, Maillefer, Baillagues, Switzerland) for each of the following sizes were used: .06-25, .06-20, and .04-25. They were divided into three groups: A = 10 new instruments (control), B = 20 used instruments operated by a high-torque motor, and C = 20 used instruments operated by a low-torque motor. Each instrument was used in 10 clinical cases (at least six molars). Cyclic fatigue testing of new and used instruments was performed with a device that allowed the instruments to rotate freely inside a stainless-steel artificial canal at 350 rpm. Data were recorded and statistically analyzed. Mean (and SD) numbers of rotation-to-breakage cycles for .06-25, .06-20, and .04-25 instruments were, respectively: group A 463 (85), 535 (101), and 613 (105); group B 229 (64), 343 (92), and 233 (84); and group C 352 (77), 454 (89), and 466 (86). Results showed that group A instruments exhibited significantly higher values of rotation-to-breakage cycles (two-sample t test, p < 0.01), when compared with groups B and C, respectively. Statistically higher resistance to cyclic fatigue was noted for group C in all sizes compared with group B (two-sample t test, p < 0.01), showing that the use of endodontic motor with lower torque values reduced cyclic fatigue of nickel-titanium rotary instruments.

Cyclic fatigue of nickel-titanium rotary instruments after clinical use with low- and high-torque endodontic motors / Gambarini, Gianluca. - In: JOURNAL OF ENDODONTICS. - ISSN 0099-2399. - 27:12(2001), pp. 772-774. [10.1097/00004770-200112000-00015]

Cyclic fatigue of nickel-titanium rotary instruments after clinical use with low- and high-torque endodontic motors

GAMBARINI, Gianluca
2001

Abstract

This study investigated the cyclic fatigue resistance of used nickel-titanium rotary instruments, clinically operated by a traditional high-torque motor and a new low-torque electric motor. Fifty instruments (ProFile, Maillefer, Baillagues, Switzerland) for each of the following sizes were used: .06-25, .06-20, and .04-25. They were divided into three groups: A = 10 new instruments (control), B = 20 used instruments operated by a high-torque motor, and C = 20 used instruments operated by a low-torque motor. Each instrument was used in 10 clinical cases (at least six molars). Cyclic fatigue testing of new and used instruments was performed with a device that allowed the instruments to rotate freely inside a stainless-steel artificial canal at 350 rpm. Data were recorded and statistically analyzed. Mean (and SD) numbers of rotation-to-breakage cycles for .06-25, .06-20, and .04-25 instruments were, respectively: group A 463 (85), 535 (101), and 613 (105); group B 229 (64), 343 (92), and 233 (84); and group C 352 (77), 454 (89), and 466 (86). Results showed that group A instruments exhibited significantly higher values of rotation-to-breakage cycles (two-sample t test, p < 0.01), when compared with groups B and C, respectively. Statistically higher resistance to cyclic fatigue was noted for group C in all sizes compared with group B (two-sample t test, p < 0.01), showing that the use of endodontic motor with lower torque values reduced cyclic fatigue of nickel-titanium rotary instruments.
2001
01 Pubblicazione su rivista::01a Articolo in rivista
Cyclic fatigue of nickel-titanium rotary instruments after clinical use with low- and high-torque endodontic motors / Gambarini, Gianluca. - In: JOURNAL OF ENDODONTICS. - ISSN 0099-2399. - 27:12(2001), pp. 772-774. [10.1097/00004770-200112000-00015]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/121927
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 22
  • Scopus 108
  • ???jsp.display-item.citation.isi??? 101
social impact