PURPOSE: To compare the visual acuity and contrast sensitivity in eyes with the AcrySof ReSTOR multifocal intraocular lens (IOL) (Alcon) and eyes with the monofocal AcrySof SA60AT IOL. SETTING: Policlinico Umberto I, Department of Ophthalmology, Rome, and private clinical practice, Rome, Italy. METHODS: One hundred eyes had phacoemulsification cataract extraction and implantation of a ReSTOR multifocal OIL in the capsular bag. Inclusion criteria were corneal astigmatism less than 1.5 diopters (D), myopia less than 4.0 D, and no associated ocular disease. A complete ophthalmic examination, including uncorrected visual acuity, best spectacle-corrected visual acuity, and contrast sensitivity, was performed 6 months postoperatively. Results were compared with those in 40 eyes with the AcrySof monofocal IOL single-piece IOL. RESULTS: In the multifocal group, 90 eyes (90%) had an uncorrected distance visual acuity of 20/25 or better (logMAR <0.10) and an uncorrected near visual acuity at 35 cm of J3 or better (logMAR 0.14). The multifocal group and monofocal group had similar distance uncorrected and best corrected visual acuities; however, the multifocal group had significantly better near uncorrected acuity. The mean contrast sensitivity values were 18.28 dB (static program) and 17.95 dB (dynamic program) in the multifocal group and 19.18 dB (static program) and 21.2 dB (dynamic program) in the monofocal group. CONCLUSIONS: The ReSTOR multifocal IOL provided a satisfactory full range of vision; 92% of the patients achieved total spectacle independence. Contrast sensitivity was lower than with the SA60AT monofocal IOL.

Visual acuity and contrast sensitivity: AcrySof ReSTOR apodized diffractive versus AcrySof SA60AT monofocal intraocular lenses / Vingolo, Enzo Maria; Pierluigi, Grenga; Luca, Iacobelli; Grenga, Roberto. - In: JOURNAL OF CATARACT AND REFRACTIVE SURGERY. - ISSN 0886-3350. - STAMPA. - 33:7(2007), pp. 1244-1247. [10.1016/j.jcrs.2007.03.052]

Visual acuity and contrast sensitivity: AcrySof ReSTOR apodized diffractive versus AcrySof SA60AT monofocal intraocular lenses

VINGOLO, Enzo Maria;GRENGA, Roberto
2007

Abstract

PURPOSE: To compare the visual acuity and contrast sensitivity in eyes with the AcrySof ReSTOR multifocal intraocular lens (IOL) (Alcon) and eyes with the monofocal AcrySof SA60AT IOL. SETTING: Policlinico Umberto I, Department of Ophthalmology, Rome, and private clinical practice, Rome, Italy. METHODS: One hundred eyes had phacoemulsification cataract extraction and implantation of a ReSTOR multifocal OIL in the capsular bag. Inclusion criteria were corneal astigmatism less than 1.5 diopters (D), myopia less than 4.0 D, and no associated ocular disease. A complete ophthalmic examination, including uncorrected visual acuity, best spectacle-corrected visual acuity, and contrast sensitivity, was performed 6 months postoperatively. Results were compared with those in 40 eyes with the AcrySof monofocal IOL single-piece IOL. RESULTS: In the multifocal group, 90 eyes (90%) had an uncorrected distance visual acuity of 20/25 or better (logMAR <0.10) and an uncorrected near visual acuity at 35 cm of J3 or better (logMAR 0.14). The multifocal group and monofocal group had similar distance uncorrected and best corrected visual acuities; however, the multifocal group had significantly better near uncorrected acuity. The mean contrast sensitivity values were 18.28 dB (static program) and 17.95 dB (dynamic program) in the multifocal group and 19.18 dB (static program) and 21.2 dB (dynamic program) in the monofocal group. CONCLUSIONS: The ReSTOR multifocal IOL provided a satisfactory full range of vision; 92% of the patients achieved total spectacle independence. Contrast sensitivity was lower than with the SA60AT monofocal IOL.
2007
01 Pubblicazione su rivista::01a Articolo in rivista
Visual acuity and contrast sensitivity: AcrySof ReSTOR apodized diffractive versus AcrySof SA60AT monofocal intraocular lenses / Vingolo, Enzo Maria; Pierluigi, Grenga; Luca, Iacobelli; Grenga, Roberto. - In: JOURNAL OF CATARACT AND REFRACTIVE SURGERY. - ISSN 0886-3350. - STAMPA. - 33:7(2007), pp. 1244-1247. [10.1016/j.jcrs.2007.03.052]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/121428
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 21
  • Scopus 119
  • ???jsp.display-item.citation.isi??? 99
social impact