The authors describe breakage of the 32 mm alumina AL2 O3 ceramic head in 3 cases of total hip replacement using two different models of prosthesis (Charnley-Müller and Mittelmeier). All three patients were of average height, weight, and activity (Brown et al., 1985; Callaghan et al., 1988), and the breakage had been caused by an accidental fall. The same mechanism of injury, in people the same age as our subjects, can cause femoral neck fractures. Reoperation was necessary in order to replace the component. The implants all appeared to be positioned correctly, and the patients had reported no symptoms. Before the trauma that caused breakage, there had been no other injuries worth noting. All patients were satisfied with their hip replacements.

"Rottura traumatica della testa protesica in ceramica in artroprotesi d'anca" / Moreschini, Oreste; Braidotti, P.. - In: GIORNALE ITALIANO DI ORTOPEDIA E TRAUMATOLOGIA. - ISSN 0390-0134. - STAMPA. - Vol. XVII, fasc 3:(1991), pp. 339-343.

"Rottura traumatica della testa protesica in ceramica in artroprotesi d'anca"

MORESCHINI, Oreste;
1991

Abstract

The authors describe breakage of the 32 mm alumina AL2 O3 ceramic head in 3 cases of total hip replacement using two different models of prosthesis (Charnley-Müller and Mittelmeier). All three patients were of average height, weight, and activity (Brown et al., 1985; Callaghan et al., 1988), and the breakage had been caused by an accidental fall. The same mechanism of injury, in people the same age as our subjects, can cause femoral neck fractures. Reoperation was necessary in order to replace the component. The implants all appeared to be positioned correctly, and the patients had reported no symptoms. Before the trauma that caused breakage, there had been no other injuries worth noting. All patients were satisfied with their hip replacements.
1991
01 Pubblicazione su rivista::01a Articolo in rivista
"Rottura traumatica della testa protesica in ceramica in artroprotesi d'anca" / Moreschini, Oreste; Braidotti, P.. - In: GIORNALE ITALIANO DI ORTOPEDIA E TRAUMATOLOGIA. - ISSN 0390-0134. - STAMPA. - Vol. XVII, fasc 3:(1991), pp. 339-343.
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/121280
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact