Objectives: To assess the image quality and diagnostic accuracy of a noise-optimized virtual monoenergetic imaging (VMI+) algorithm compared with standard virtual monoenergetic imaging (VMI) and linearly-blended (M_0.6) reconstructions for the detection of hypervascular liver lesions in dual-energy CT (DECT). Methods: Thirty patients who underwent clinical liver MRI were prospectively enrolled. Within 60 days of MRI, arterial phase DECT images were acquired on a third-generation dual-source CT and reconstructed with M_0.6, VMI and VMI+ algorithms from 40 to 100 keV in 5-keV intervals. Liver parenchyma and lesion contrast-to-noise-ratios (CNR) were calculated. Two radiologists assessed image quality. Lesion sensitivity, specificity and area under the receiver operating characteristic curves (AUCs) were calculated for the three algorithms with MRI as the reference standard. Results: VMI+ datasets from 40 to 60 keV provided the highest liver parenchyma and lesion CNR (p ≤0.021); 50 keV VMI+ provided the highest subjective image quality (4.40±0.54), significantly higher compared to VMI and M_0.6 (all p <0.001), and the best diagnostic accuracy in < 1-cm diameter lesions (AUC=0.833 vs. 0.777 and 0.749, respectively; p ≤0.003). Conclusions: 50-keV VMI+ provides superior image quality and diagnostic accuracy for the detection of hypervascular liver lesions with a diameter < 1cm compared to VMI or M_0.6 reconstructions. Key Points: • Low-keV VMI+ are characterized by higher contrast resulting from maximum iodine attenuation. • VMI+ provides superior image quality compared with VMI or M_0.6. • 50-keV_VMI+ provides higher accuracy for the detection of hypervascular liver lesions < 1cm.

A noise-optimized virtual monoenergetic reconstruction algorithm improves the diagnostic accuracy of late hepatic arterial phase dual-energy CT for the detection of hypervascular liver lesions / De Cecco, Carlo N.; Caruso, Damiano; Joseph Schoepf, U.; DE SANTIS, Domenico; Muscogiuri, Giuseppe; Albrecht, Moritz H.; Meinel, Felix G.; Wichmann, Julian L.; Burchett, Philip F.; Varga-Szemes, Akos; Sheafor, Douglas H.; Hardie, Andrew D.. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - 28:8(2018), pp. 3393-3404. [10.1007/s00330-018-5313-6]

A noise-optimized virtual monoenergetic reconstruction algorithm improves the diagnostic accuracy of late hepatic arterial phase dual-energy CT for the detection of hypervascular liver lesions

Carlo N. De Cecco;Damiano Caruso;Domenico De Santis;Giuseppe Muscogiuri;
2018

Abstract

Objectives: To assess the image quality and diagnostic accuracy of a noise-optimized virtual monoenergetic imaging (VMI+) algorithm compared with standard virtual monoenergetic imaging (VMI) and linearly-blended (M_0.6) reconstructions for the detection of hypervascular liver lesions in dual-energy CT (DECT). Methods: Thirty patients who underwent clinical liver MRI were prospectively enrolled. Within 60 days of MRI, arterial phase DECT images were acquired on a third-generation dual-source CT and reconstructed with M_0.6, VMI and VMI+ algorithms from 40 to 100 keV in 5-keV intervals. Liver parenchyma and lesion contrast-to-noise-ratios (CNR) were calculated. Two radiologists assessed image quality. Lesion sensitivity, specificity and area under the receiver operating characteristic curves (AUCs) were calculated for the three algorithms with MRI as the reference standard. Results: VMI+ datasets from 40 to 60 keV provided the highest liver parenchyma and lesion CNR (p ≤0.021); 50 keV VMI+ provided the highest subjective image quality (4.40±0.54), significantly higher compared to VMI and M_0.6 (all p <0.001), and the best diagnostic accuracy in < 1-cm diameter lesions (AUC=0.833 vs. 0.777 and 0.749, respectively; p ≤0.003). Conclusions: 50-keV VMI+ provides superior image quality and diagnostic accuracy for the detection of hypervascular liver lesions with a diameter < 1cm compared to VMI or M_0.6 reconstructions. Key Points: • Low-keV VMI+ are characterized by higher contrast resulting from maximum iodine attenuation. • VMI+ provides superior image quality compared with VMI or M_0.6. • 50-keV_VMI+ provides higher accuracy for the detection of hypervascular liver lesions < 1cm.
2018
diagnostic accuracy; dual-energy CT; hypervascular liver lesions; liver imaging; virtual monoenergetic images; radiology, nuclear medicine and imaging
01 Pubblicazione su rivista::01a Articolo in rivista
A noise-optimized virtual monoenergetic reconstruction algorithm improves the diagnostic accuracy of late hepatic arterial phase dual-energy CT for the detection of hypervascular liver lesions / De Cecco, Carlo N.; Caruso, Damiano; Joseph Schoepf, U.; DE SANTIS, Domenico; Muscogiuri, Giuseppe; Albrecht, Moritz H.; Meinel, Felix G.; Wichmann, Julian L.; Burchett, Philip F.; Varga-Szemes, Akos; Sheafor, Douglas H.; Hardie, Andrew D.. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - 28:8(2018), pp. 3393-3404. [10.1007/s00330-018-5313-6]
File allegati a questo prodotto
File Dimensione Formato  
DeCecco_Noise-optimized_2018.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 2.13 MB
Formato Adobe PDF
2.13 MB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1166300
Citazioni
  • ???jsp.display-item.citation.pmc??? 14
  • Scopus 52
  • ???jsp.display-item.citation.isi??? 53
social impact