Objectives: To define optimal window settings for displaying virtual monoenergetic images (VMI) of dual-energy CT pulmonary angiography (DE-CTPA). Methods: Forty-five patients who underwent clinically-indicated third-generation dual-source DE-CTPA were retrospectively evaluated. Standard linearly-blended (M_0.6), 70-keV traditional VMI (M70), and 40-keV noise-optimised VMI (M40+) reconstructions were analysed. For M70 and M40+ datasets, the subjectively best window setting (width and level, B-W/L) was independently determined by two observers and subsequently related with pulmonary artery attenuation to calculate separate optimised values (O-W/L) using linear regression. Subjective evaluation of image quality (IQ) between W/L settings were assessed by two additional readers. Repeated measures of variance were performed to compare W/L settings and IQ indices between M_0.6, M70, and M40+. Results: B-W/L and O-W/L for M70 were 460/140 and 450/140, and were 1100/380 and 1070/380 for M40+, respectively, differing from standard DE-CTPA W/L settings (450/100). Highest subjective scores were observed for M40+ regarding vascular contrast, embolism demarcation, and overall IQ (all pOpenSPiltSPi0.001). Conclusions: Application of O-W/L settings is beneficial to optimise subjective IQ of VMI reconstructions of DE-CTPA. A width slightly less than two times the pulmonary trunk attenuation and a level approximately of overall pulmonary vessel attenuation are recommended. Key Points: • Application of standard window settings for VMI results in inferior image perception. • No significant differences between B-W/L and O-W/L for M70/M40+ were observed. • O-W/L for M70 were 450/140 and were 1070/380 for M40+. • Improved subjective IQ characteristics were observed for VMI displayed with O-W/L.

Optimisation of window settings for traditional and noise-optimised virtual monoenergetic imaging in dual-energy computed tomography pulmonary angiography / Tommaso, D’Angelo; Bucher, Andreas M.; Lenga, Lukas; Arendt, Christophe T.; Peterke, Julia L.; Caruso, Damiano; Mazziotti, Silvio; Blandino, Alfredo; Ascenti, Giorgio; Othman, Ahmed E.; Martin, Simon S.; Leithner, Doris; Vogl, Thomas J.; Wichmann, Julian L.. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - 28:4(2018), pp. 1393-1401. [10.1007/s00330-017-5059-6]

Optimisation of window settings for traditional and noise-optimised virtual monoenergetic imaging in dual-energy computed tomography pulmonary angiography

Tommaso D’Angelo;Damiano Caruso;
2018

Abstract

Objectives: To define optimal window settings for displaying virtual monoenergetic images (VMI) of dual-energy CT pulmonary angiography (DE-CTPA). Methods: Forty-five patients who underwent clinically-indicated third-generation dual-source DE-CTPA were retrospectively evaluated. Standard linearly-blended (M_0.6), 70-keV traditional VMI (M70), and 40-keV noise-optimised VMI (M40+) reconstructions were analysed. For M70 and M40+ datasets, the subjectively best window setting (width and level, B-W/L) was independently determined by two observers and subsequently related with pulmonary artery attenuation to calculate separate optimised values (O-W/L) using linear regression. Subjective evaluation of image quality (IQ) between W/L settings were assessed by two additional readers. Repeated measures of variance were performed to compare W/L settings and IQ indices between M_0.6, M70, and M40+. Results: B-W/L and O-W/L for M70 were 460/140 and 450/140, and were 1100/380 and 1070/380 for M40+, respectively, differing from standard DE-CTPA W/L settings (450/100). Highest subjective scores were observed for M40+ regarding vascular contrast, embolism demarcation, and overall IQ (all pOpenSPiltSPi0.001). Conclusions: Application of O-W/L settings is beneficial to optimise subjective IQ of VMI reconstructions of DE-CTPA. A width slightly less than two times the pulmonary trunk attenuation and a level approximately of overall pulmonary vessel attenuation are recommended. Key Points: • Application of standard window settings for VMI results in inferior image perception. • No significant differences between B-W/L and O-W/L for M70/M40+ were observed. • O-W/L for M70 were 450/140 and were 1070/380 for M40+. • Improved subjective IQ characteristics were observed for VMI displayed with O-W/L.
2018
computed tomography pulmonary angiography; dual-energy CT; pulmonary embolism; window level; window width; computed tomography angiography
01 Pubblicazione su rivista::01a Articolo in rivista
Optimisation of window settings for traditional and noise-optimised virtual monoenergetic imaging in dual-energy computed tomography pulmonary angiography / Tommaso, D’Angelo; Bucher, Andreas M.; Lenga, Lukas; Arendt, Christophe T.; Peterke, Julia L.; Caruso, Damiano; Mazziotti, Silvio; Blandino, Alfredo; Ascenti, Giorgio; Othman, Ahmed E.; Martin, Simon S.; Leithner, Doris; Vogl, Thomas J.; Wichmann, Julian L.. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - 28:4(2018), pp. 1393-1401. [10.1007/s00330-017-5059-6]
File allegati a questo prodotto
File Dimensione Formato  
DAngelo_Optimisation_2018.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 787.05 kB
Formato Adobe PDF
787.05 kB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1166263
Citazioni
  • ???jsp.display-item.citation.pmc??? 11
  • Scopus 30
  • ???jsp.display-item.citation.isi??? 25
social impact