Introduction: The aim of the present study was to investigate the impact of ambient temperature on the life span of nickel-titanium (NiTi) files. Methods: Cyclic fatigue testing of NiTi instruments was performed in a stainless steel artificial canal. During the experiment, 4 different temperatures (0°C, 20°C, 35°C, and 39°C) were used in a thermostatic bath controlled using electronic and infrared controls. Eighty R25 Reciproc Blue (VDW, Munich, Germany) and 80 R25 Reciproc (VDW) instruments were reciprocated in the artificial canal until fracture occurred, and the time to fracture and the length of the fractured fragment were recorded. The phase transformation temperature for 3 instruments of each type was analyzed by differential scanning calorimetry. One-way analysis of variance and Bonferroni tests were used to statistically analyze the data at a 5% significance level (P <.05). Results: Reciproc Blue instruments were significantly more resistant to cyclic fatigue than Reciproc at all the temperatures tested (P <.05). Fatigue resistance was proportionally higher as the environmental temperature decreased (0°C > 20°C > 35°C > 39°C). No significant difference was registered for the length of the fractured fragment of both instruments tested at all temperatures tested (P >.05). Conclusions: Blue alloy significantly increases the cyclic fatigue resistance of Reciproc files compared with M-Wire alloy. Temperature significantly affects the life span of NiTi files. When the ambient temperature increases between 0°C and 35°C, the fatigue resistance of the files statistically decreases.

Cyclic fatigue of Reciproc and Reciproc Blue nickel-titanium reciprocating files at different environmental temperatures / Plotino, Gianluca; Grande, Nicola M.; Testarelli, Luca; Gambarini, Gianluca; Castagnola, Raffaella; Rossetti, Andrea; Özyürek, Taha; Cordaro, Massimo; Fortunato, Leonzio. - In: JOURNAL OF ENDODONTICS. - ISSN 0099-2399. - STAMPA. - 44:10(2018), pp. 1549-1552. [10.1016/j.joen.2018.06.006]

Cyclic fatigue of Reciproc and Reciproc Blue nickel-titanium reciprocating files at different environmental temperatures

Plotino, Gianluca
Primo
;
Grande, Nicola M.
Secondo
;
Testarelli, Luca
Methodology
;
Gambarini, Gianluca;
2018

Abstract

Introduction: The aim of the present study was to investigate the impact of ambient temperature on the life span of nickel-titanium (NiTi) files. Methods: Cyclic fatigue testing of NiTi instruments was performed in a stainless steel artificial canal. During the experiment, 4 different temperatures (0°C, 20°C, 35°C, and 39°C) were used in a thermostatic bath controlled using electronic and infrared controls. Eighty R25 Reciproc Blue (VDW, Munich, Germany) and 80 R25 Reciproc (VDW) instruments were reciprocated in the artificial canal until fracture occurred, and the time to fracture and the length of the fractured fragment were recorded. The phase transformation temperature for 3 instruments of each type was analyzed by differential scanning calorimetry. One-way analysis of variance and Bonferroni tests were used to statistically analyze the data at a 5% significance level (P <.05). Results: Reciproc Blue instruments were significantly more resistant to cyclic fatigue than Reciproc at all the temperatures tested (P <.05). Fatigue resistance was proportionally higher as the environmental temperature decreased (0°C > 20°C > 35°C > 39°C). No significant difference was registered for the length of the fractured fragment of both instruments tested at all temperatures tested (P >.05). Conclusions: Blue alloy significantly increases the cyclic fatigue resistance of Reciproc files compared with M-Wire alloy. Temperature significantly affects the life span of NiTi files. When the ambient temperature increases between 0°C and 35°C, the fatigue resistance of the files statistically decreases.
2018
ambient temperature; cyclic fatigue resistance; heat treatment; instruments; nickel-titanium; dentistry (all)
01 Pubblicazione su rivista::01a Articolo in rivista
Cyclic fatigue of Reciproc and Reciproc Blue nickel-titanium reciprocating files at different environmental temperatures / Plotino, Gianluca; Grande, Nicola M.; Testarelli, Luca; Gambarini, Gianluca; Castagnola, Raffaella; Rossetti, Andrea; Özyürek, Taha; Cordaro, Massimo; Fortunato, Leonzio. - In: JOURNAL OF ENDODONTICS. - ISSN 0099-2399. - STAMPA. - 44:10(2018), pp. 1549-1552. [10.1016/j.joen.2018.06.006]
File allegati a questo prodotto
File Dimensione Formato  
Plotino_Cyclic_2018.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 299.63 kB
Formato Adobe PDF
299.63 kB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1154597
Citazioni
  • ???jsp.display-item.citation.pmc??? 17
  • Scopus 41
  • ???jsp.display-item.citation.isi??? 33
social impact