We are going to describe the design and the electrical performance of the horn-coupled lumped element kinetic inductance detectors (LEKIDs) for the OLIMPO experiment. OLIMPO is a balloon-borne mission, devoted to the study of the largest structures in the Universe, by detecting the Sunyaev-Zel'dovich effect of the Cosmic Microwave Background (CMB) photons crossing clusters of galaxies. The multi-band focal planes (centered at 150, 200, 350 and 480 GHz), large aperture telescope (2.6 m), photometric and spectrometer configuration (OLIMPO is equipped with a plug-in differential Fourier transform spectrometer) make OLIMPO able to characaterise all the components of the Sunyaev-Zel'dovich effect with unprecedented precision. The design of the LEKIDs has been optimised to cope with a wide range of optical loadings (corresponding to both photometric and spectrometric configurations). The electrical characterisation of the LEKID arrays was performed in the OLIMPO cryostat, at a temperature around 300 mK and under a constant optical load lower than 500 fW. The readout electronics is a customised ROACH-2 coupled to a MUSIC DAC/ADC board. The averaged responsivities for the arrays, measured from the phase readout, are 1.18x1011rad/W for the 150 GHz array, 4.10x1010rad/W for the 200 GHz array, 1.09x1012rad/W for the 350 GHz array, and 4.41x1011rad/W for the 480 GHz array. Therefore, measuring the spectral noise density for each pixels of the arrays, whose average values are 2.5x10-5rad/Hz(1/2)for the 150 GHz array, 3.6x10-5rad/Hz(1/2)for the 200 GHz array, 5.1x10-5rad/Hz(1/2)for the 350 GHz array, and 1.9x10-5rad/Hz(1/2)for the 480 GHz array, we obtained the values of the averaged electrical noise equivalent power (NEP) over all the arrays: 2x10-16W/Hz(1/2), 9x10-16W/Hz(1/2), 5x10-17W/Hz(1/2), and 4x10-17W/Hz(1/2)respectively for the four OLIMPO arrays.

Design and Electrical Performance of the Kinetic Inductance Detectors of the OLIMPO Experiment / Paiella, A.; Battistelli, E. S.; Columbro, F.; Coppolecchia, A.; D'Alessandro, G.; De Bernardis, P.; Lamagna, L.; Masi, S.; Piacentini, F.; Castellano, M. G.; Colantoni, I.; Gordon, S.; Mauskopf, P.. - 2018-:(2018), pp. 1-3. (Intervento presentato al convegno 16th International Superconductive Electronics Conference, ISEC 2017 tenutosi a ita) [10.1109/ISEC.2017.8314223].

Design and Electrical Performance of the Kinetic Inductance Detectors of the OLIMPO Experiment

Paiella, A.
;
Battistelli, E. S.;Columbro, F.;Coppolecchia, A.;D'Alessandro, G.;De Bernardis, P.;Lamagna, L.;Masi, S.;Piacentini, F.;Castellano, M. G.;Colantoni, I.;Mauskopf, P.
2018

Abstract

We are going to describe the design and the electrical performance of the horn-coupled lumped element kinetic inductance detectors (LEKIDs) for the OLIMPO experiment. OLIMPO is a balloon-borne mission, devoted to the study of the largest structures in the Universe, by detecting the Sunyaev-Zel'dovich effect of the Cosmic Microwave Background (CMB) photons crossing clusters of galaxies. The multi-band focal planes (centered at 150, 200, 350 and 480 GHz), large aperture telescope (2.6 m), photometric and spectrometer configuration (OLIMPO is equipped with a plug-in differential Fourier transform spectrometer) make OLIMPO able to characaterise all the components of the Sunyaev-Zel'dovich effect with unprecedented precision. The design of the LEKIDs has been optimised to cope with a wide range of optical loadings (corresponding to both photometric and spectrometric configurations). The electrical characterisation of the LEKID arrays was performed in the OLIMPO cryostat, at a temperature around 300 mK and under a constant optical load lower than 500 fW. The readout electronics is a customised ROACH-2 coupled to a MUSIC DAC/ADC board. The averaged responsivities for the arrays, measured from the phase readout, are 1.18x1011rad/W for the 150 GHz array, 4.10x1010rad/W for the 200 GHz array, 1.09x1012rad/W for the 350 GHz array, and 4.41x1011rad/W for the 480 GHz array. Therefore, measuring the spectral noise density for each pixels of the arrays, whose average values are 2.5x10-5rad/Hz(1/2)for the 150 GHz array, 3.6x10-5rad/Hz(1/2)for the 200 GHz array, 5.1x10-5rad/Hz(1/2)for the 350 GHz array, and 1.9x10-5rad/Hz(1/2)for the 480 GHz array, we obtained the values of the averaged electrical noise equivalent power (NEP) over all the arrays: 2x10-16W/Hz(1/2), 9x10-16W/Hz(1/2), 5x10-17W/Hz(1/2), and 4x10-17W/Hz(1/2)respectively for the four OLIMPO arrays.
2018
16th International Superconductive Electronics Conference, ISEC 2017
CMB radiation; kinetic inductance detectors; OLIMPO experiment; SZ effect; Electrical and Electronic Engineering; Electronic, Optical and Magnetic Materials; Atomic and Molecular Physics, and Optics
04 Pubblicazione in atti di convegno::04b Atto di convegno in volume
Design and Electrical Performance of the Kinetic Inductance Detectors of the OLIMPO Experiment / Paiella, A.; Battistelli, E. S.; Columbro, F.; Coppolecchia, A.; D'Alessandro, G.; De Bernardis, P.; Lamagna, L.; Masi, S.; Piacentini, F.; Castellano, M. G.; Colantoni, I.; Gordon, S.; Mauskopf, P.. - 2018-:(2018), pp. 1-3. (Intervento presentato al convegno 16th International Superconductive Electronics Conference, ISEC 2017 tenutosi a ita) [10.1109/ISEC.2017.8314223].
File allegati a questo prodotto
File Dimensione Formato  
Paiella_Design_2017.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 1.6 MB
Formato Adobe PDF
1.6 MB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1154432
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 8
social impact