We have developed a high-performance liquid chromatographic-UV-Vis-diode-array detection (HPLC-DAD) method for the determination of ethyl-p-hydroxybenzoate, a hydrolytic degradation product of the synthetic protease inhibitor, gabexate-mesilate ethyl-p-(6-guanidinohexanoyloxy) benzoate methanesulfonate (GM) (FOY) in sow pancreatic juice. Methyl-p-hydroxybenzoate (I) was used as the internal standard. The pancreatic juice was deproteinised by acetonitrile and the analytes were chromatographed on a reversed-phase C18 LC column using the gradient elution method. The mobile phase consisted of a solution of 0.017 M orthophosphoric acid and another solution of acetonitrile-water (80:20, v/v). The wavelength of detection was 237 nm. The limit of quantification of the method was 0.20 μM at a 9:1 signal-to-noise ratio. The overall intra- and inter-day accuracy (relative error, RE) ranged from 14.2 to 8.3% and from 13.3 to 9.8, respectively. The overall intra- and inter-day precision (relative standard deviation, RSD) ranged from 7.6 to 2.62% and from 6.7 to 3.1%, respectively. The method proved to be sensitive, specific, accurate and precise and was successfully used to determine the ethyl-p-hydroxybenzoate (II) in sow pancreatic juice. © 2001 Elsevier Science B.V.

Determination of ethyl-p-hydroxybenzoate in sow pancreatic juice by reversed-phase high-performance liquid chromatography / R., Di Giovannandrea; L., Diana; E., Ferretti; G., Foglietta; Caronna, Roberto; G., Severini. - In: JOURNAL OF CHROMATOGRAPHY B. BIOMEDICAL SCIENCES AND APPLICATIONS. - ISSN 1387-2273. - STAMPA. - 751:2(2001), pp. 365-369. [10.1016/s0378-4347(00)00477-1]

Determination of ethyl-p-hydroxybenzoate in sow pancreatic juice by reversed-phase high-performance liquid chromatography

CARONNA, Roberto;
2001

Abstract

We have developed a high-performance liquid chromatographic-UV-Vis-diode-array detection (HPLC-DAD) method for the determination of ethyl-p-hydroxybenzoate, a hydrolytic degradation product of the synthetic protease inhibitor, gabexate-mesilate ethyl-p-(6-guanidinohexanoyloxy) benzoate methanesulfonate (GM) (FOY) in sow pancreatic juice. Methyl-p-hydroxybenzoate (I) was used as the internal standard. The pancreatic juice was deproteinised by acetonitrile and the analytes were chromatographed on a reversed-phase C18 LC column using the gradient elution method. The mobile phase consisted of a solution of 0.017 M orthophosphoric acid and another solution of acetonitrile-water (80:20, v/v). The wavelength of detection was 237 nm. The limit of quantification of the method was 0.20 μM at a 9:1 signal-to-noise ratio. The overall intra- and inter-day accuracy (relative error, RE) ranged from 14.2 to 8.3% and from 13.3 to 9.8, respectively. The overall intra- and inter-day precision (relative standard deviation, RSD) ranged from 7.6 to 2.62% and from 6.7 to 3.1%, respectively. The method proved to be sensitive, specific, accurate and precise and was successfully used to determine the ethyl-p-hydroxybenzoate (II) in sow pancreatic juice. © 2001 Elsevier Science B.V.
2001
ethyl-p-hydroxybenzoate
01 Pubblicazione su rivista::01a Articolo in rivista
Determination of ethyl-p-hydroxybenzoate in sow pancreatic juice by reversed-phase high-performance liquid chromatography / R., Di Giovannandrea; L., Diana; E., Ferretti; G., Foglietta; Caronna, Roberto; G., Severini. - In: JOURNAL OF CHROMATOGRAPHY B. BIOMEDICAL SCIENCES AND APPLICATIONS. - ISSN 1387-2273. - STAMPA. - 751:2(2001), pp. 365-369. [10.1016/s0378-4347(00)00477-1]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/113869
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 10
  • ???jsp.display-item.citation.isi??? 9
social impact