Introduction The aim of this study was to compare the accuracy of 2 electronic apex locators (EALs), Apex ID (SybronEndo, Glendora, CA) and Root ZX (J. Morita, Tokyo, Japan), by means of micro–computed tomographic (micro-CT) imaging and to determine anatomic variations that may affect their accuracy. Methods The root canal length (RCL) and working length (WL) of 33 single-rooted premolars were measured using a visual method, 3-dimensional micro-CT reconstructions, and 2 different EALs. Two different measurements were recorded for each EAL: at the “APEX/0.0” mark and at the “0.5” mark. The WL was determined using 2 different methods: method 1: at the “0.5 mark” of the apex locator and method 2: subtracting 0.5 mm from the “APEX/0.0” mark. The precision of measurements was compared with those recorded by micro-CT imaging. Apical foramen (AF) position and diameter, apical constriction (AC) diameter, distance between the AC and the AF, and the presence/absence of accessory canals were recorded from the micro-CT scans, and their correlation to the accuracy of EALs was determined. Results There was no statistically significant difference in the RCL measurements by any of the different methods. There was a statistically significant difference in the WL recorded by micro-CT imaging compared with those by the visual method and at the “APEX/0.0 mark” − 0.5 mm (P = .031). There was no difference in the measurements acquired by any of the EALs. The “APEX/0.0 mark” − 0.5 mm was less accurate than the “0.5” mark. However, the results were not statistically significant (P > .05). The position of the AF and the AC–AF distance affected the accuracy of the RCL (P = .003) and the “0.5” mark (P = .013). Conclusions Root ZX and Apex ID are equally precise in determining the RCL and WL. The “0.5” mark can be used to determine the WL with high precision. Some anatomic variations may influence the accuracy of EALs.

The use of micro–computed tomography to determine the accuracy of 2 electronic apex locators and anatomic variations affecting their precision / Piasecki, Lucila; Carneiro, Everdan; da Silva Neto, Ulisses Xavier; Westphalen, Vânia Portela Ditzel; Brandão, Christian Giampietro; Gambarini, Gianluca; Azim, Adham A.. - In: JOURNAL OF ENDODONTICS. - ISSN 0099-2399. - STAMPA. - 42:8(2016), pp. 1263-1267. [10.1016/j.joen.2016.04.024]

The use of micro–computed tomography to determine the accuracy of 2 electronic apex locators and anatomic variations affecting their precision

Gambarini, Gianluca
Penultimo
;
2016

Abstract

Introduction The aim of this study was to compare the accuracy of 2 electronic apex locators (EALs), Apex ID (SybronEndo, Glendora, CA) and Root ZX (J. Morita, Tokyo, Japan), by means of micro–computed tomographic (micro-CT) imaging and to determine anatomic variations that may affect their accuracy. Methods The root canal length (RCL) and working length (WL) of 33 single-rooted premolars were measured using a visual method, 3-dimensional micro-CT reconstructions, and 2 different EALs. Two different measurements were recorded for each EAL: at the “APEX/0.0” mark and at the “0.5” mark. The WL was determined using 2 different methods: method 1: at the “0.5 mark” of the apex locator and method 2: subtracting 0.5 mm from the “APEX/0.0” mark. The precision of measurements was compared with those recorded by micro-CT imaging. Apical foramen (AF) position and diameter, apical constriction (AC) diameter, distance between the AC and the AF, and the presence/absence of accessory canals were recorded from the micro-CT scans, and their correlation to the accuracy of EALs was determined. Results There was no statistically significant difference in the RCL measurements by any of the different methods. There was a statistically significant difference in the WL recorded by micro-CT imaging compared with those by the visual method and at the “APEX/0.0 mark” − 0.5 mm (P = .031). There was no difference in the measurements acquired by any of the EALs. The “APEX/0.0 mark” − 0.5 mm was less accurate than the “0.5” mark. However, the results were not statistically significant (P > .05). The position of the AF and the AC–AF distance affected the accuracy of the RCL (P = .003) and the “0.5” mark (P = .013). Conclusions Root ZX and Apex ID are equally precise in determining the RCL and WL. The “0.5” mark can be used to determine the WL with high precision. Some anatomic variations may influence the accuracy of EALs.
2016
apex ID; apical limit; electronic apex locator; micro–computed tomographic imaging; root ZX; working length; bicuspid; dental pulp cavity; electronics, medical; humans; in vitro techniques; odontometry; root canal preparation; tooth apex; x-ray microtomography; dentistry (all)
01 Pubblicazione su rivista::01a Articolo in rivista
The use of micro–computed tomography to determine the accuracy of 2 electronic apex locators and anatomic variations affecting their precision / Piasecki, Lucila; Carneiro, Everdan; da Silva Neto, Ulisses Xavier; Westphalen, Vânia Portela Ditzel; Brandão, Christian Giampietro; Gambarini, Gianluca; Azim, Adham A.. - In: JOURNAL OF ENDODONTICS. - ISSN 0099-2399. - STAMPA. - 42:8(2016), pp. 1263-1267. [10.1016/j.joen.2016.04.024]
File allegati a questo prodotto
File Dimensione Formato  
Piasecki_Use-of-micro_2016_1.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 328.82 kB
Formato Adobe PDF
328.82 kB Adobe PDF   Contatta l'autore
Piasecki_Use-of-micro_2016_2.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 176.74 kB
Formato Adobe PDF
176.74 kB Adobe PDF   Contatta l'autore
Piasecki_Use-of-micro_2016_3.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 26.26 kB
Formato Adobe PDF
26.26 kB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1112346
Citazioni
  • ???jsp.display-item.citation.pmc??? 6
  • Scopus 44
  • ???jsp.display-item.citation.isi??? 30
social impact