Objective: The purpose of this study was to assess the inter- and intra-assessor reliability of the cervical spine device (Formetric, DIERS International GmbH, Schlangenbad, Germany) in measuring cervical range of motion. Methods: The cervical spine device was used to measure the cervical range of motion of 65 asymptomatic participants. Flexion-extension, right and left rotation, and right and left lateral flexion were analyzed. Two different assessors performed the measurements on the same day to estimate inter-assessor reliability and 2 days later to examine intra-assessor reliability. Intra-assessor and inter-assessor reliability was assessed using the intraclass correlation coefficient (ICC). The standard error of measurement (SEM) and the smallest detectable difference (SDD) were also estimated. Results: Inter-assessor reliability ICCs for flexion + extension and total lateral flexion movements were >0.90. The ICCs for rotation movements and for left lateral flexion were >0.70. The ICCs for flexion (0.64), extension (0.58), and right lateral flexion (0.56) indicated moderate correlation. Mean SEMs ranged from 2.28° (SDD = 6.31°) for left rotation to 8.08° (SDD = 22.38°) for total rotation. As for intra-assessor test-retest reliability, all ICCs were >0.70. Mean SEMs ranged from 3.14° (SDD = 8.70°) for total lateral flexion to 7.50° (SDD = 20.77°) for extension. Conclusion: Both inter- and intra-observer reproducibility correlation values are moderate to high for measurements obtained using the cervical spine device.

Reliability of the cervical spine device for the assessment of cervical spine range of motion in asymptomatic participants / Mangone, Massimiliano; Bernetti, Andrea; Germanotta, Marco; Di Sipio, Enrica; Razzano, Cristina; Ioppolo, Francesco; Santilli, Valter; Venditto, Teresa; Paoloni, Marco. - In: JOURNAL OF MANIPULATIVE AND PHYSIOLOGICAL THERAPEUTICS. - ISSN 0161-4754. - STAMPA. - 41:4(2018), pp. 342-349. [10.1016/j.jmpt.2018.01.003]

Reliability of the cervical spine device for the assessment of cervical spine range of motion in asymptomatic participants

Mangone, Massimiliano;Bernetti, Andrea;Germanotta, Marco;Razzano, Cristina;Ioppolo, Francesco;Santilli, Valter;Venditto, Teresa
;
Paoloni, Marco
2018

Abstract

Objective: The purpose of this study was to assess the inter- and intra-assessor reliability of the cervical spine device (Formetric, DIERS International GmbH, Schlangenbad, Germany) in measuring cervical range of motion. Methods: The cervical spine device was used to measure the cervical range of motion of 65 asymptomatic participants. Flexion-extension, right and left rotation, and right and left lateral flexion were analyzed. Two different assessors performed the measurements on the same day to estimate inter-assessor reliability and 2 days later to examine intra-assessor reliability. Intra-assessor and inter-assessor reliability was assessed using the intraclass correlation coefficient (ICC). The standard error of measurement (SEM) and the smallest detectable difference (SDD) were also estimated. Results: Inter-assessor reliability ICCs for flexion + extension and total lateral flexion movements were >0.90. The ICCs for rotation movements and for left lateral flexion were >0.70. The ICCs for flexion (0.64), extension (0.58), and right lateral flexion (0.56) indicated moderate correlation. Mean SEMs ranged from 2.28° (SDD = 6.31°) for left rotation to 8.08° (SDD = 22.38°) for total rotation. As for intra-assessor test-retest reliability, all ICCs were >0.70. Mean SEMs ranged from 3.14° (SDD = 8.70°) for total lateral flexion to 7.50° (SDD = 20.77°) for extension. Conclusion: Both inter- and intra-observer reproducibility correlation values are moderate to high for measurements obtained using the cervical spine device.
2018
range of motion; spine; tendinopathy
01 Pubblicazione su rivista::01a Articolo in rivista
Reliability of the cervical spine device for the assessment of cervical spine range of motion in asymptomatic participants / Mangone, Massimiliano; Bernetti, Andrea; Germanotta, Marco; Di Sipio, Enrica; Razzano, Cristina; Ioppolo, Francesco; Santilli, Valter; Venditto, Teresa; Paoloni, Marco. - In: JOURNAL OF MANIPULATIVE AND PHYSIOLOGICAL THERAPEUTICS. - ISSN 0161-4754. - STAMPA. - 41:4(2018), pp. 342-349. [10.1016/j.jmpt.2018.01.003]
File allegati a questo prodotto
File Dimensione Formato  
Mangone_Rehability_2018.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 288.49 kB
Formato Adobe PDF
288.49 kB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1109966
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 7
social impact