After a brief review about the principal risks of infection caused by the application of dialysis, the Authors describe the results relative to the control of the bacteriological contamination present in a dialysis unit of a large Roman hospital. The results showed a general compliance with the guidelines. However the older models of the apparatus were found to be more contaminated than the more up to date models and the contamination risk was higher in correspondence with connections of the apparatus. Furthermore the presence of opportunistic microrganisms such as P aeruginosa suggests the need for their substitution.
Contaminazione batterica in un impianto di dialisi / Fabiani, M; Montacutelli, R; Raponi, M; Orsi, Giovanni Battista. - In: ANNALI DI IGIENE MEDICINA PREVENTIVA E DI COMUNITÀ. - ISSN 1120-9135. - STAMPA. - 11:2(1999), pp. 107-116.
Contaminazione batterica in un impianto di dialisi
ORSI, Giovanni Battista
1999
Abstract
After a brief review about the principal risks of infection caused by the application of dialysis, the Authors describe the results relative to the control of the bacteriological contamination present in a dialysis unit of a large Roman hospital. The results showed a general compliance with the guidelines. However the older models of the apparatus were found to be more contaminated than the more up to date models and the contamination risk was higher in correspondence with connections of the apparatus. Furthermore the presence of opportunistic microrganisms such as P aeruginosa suggests the need for their substitution.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.