This work presents the first study of x-ray phase contrast imaging based on a simple implementation of the edge illumination method (EIXPCi) in the field of mammography with synchrotron radiation. A simplified EIXPCi set-up was utilized to study a possible application in mammography at clinical doses. Moreover, through a novel algorithm capable of separating and quantifying absorption and phase perturbations of images acquired in EIXPCi modality, it is possible to extract quantitative information on breast images, allowing an accurate tissue identification. The study was carried out at the SYRMEP beamline of Elettra synchrotron radiation facility (Trieste, Italy), where a mastectomy specimen was investigated with the EIXPCi technique. The sample was exposed at three different energies suitable for mammography with synchrotron radiation in order to test the validity of the novel algorithm in extracting values of linear attenuation coefficients integrated over the sample thickness. It is demonstrated that the quantitative data are in good agreement with the theoretical values of linear attenuation coefficients calculated on the hypothesis of the breast with a given composition. The results are promising and encourage the current efforts to apply the method in mammography with synchrotron radiation.

A simplified edge illumination set-up for quantitative phase contrast mammography with synchrotron radiation at clinical doses / Longo, Mariaconcetta; Rigon, Luigi; Lopez, Frances C. M.; Chen, Rongchang; Dreossi, Diego; Zanconati, Fabrizio; Longo, Renata. - In: PHYSICS IN MEDICINE AND BIOLOGY. - ISSN 0031-9155. - ELETTRONICO. - 60:3(2015), pp. N21-N34. [10.1088/0031-9155/60/3/N21]

A simplified edge illumination set-up for quantitative phase contrast mammography with synchrotron radiation at clinical doses

Longo, Mariaconcetta;
2015

Abstract

This work presents the first study of x-ray phase contrast imaging based on a simple implementation of the edge illumination method (EIXPCi) in the field of mammography with synchrotron radiation. A simplified EIXPCi set-up was utilized to study a possible application in mammography at clinical doses. Moreover, through a novel algorithm capable of separating and quantifying absorption and phase perturbations of images acquired in EIXPCi modality, it is possible to extract quantitative information on breast images, allowing an accurate tissue identification. The study was carried out at the SYRMEP beamline of Elettra synchrotron radiation facility (Trieste, Italy), where a mastectomy specimen was investigated with the EIXPCi technique. The sample was exposed at three different energies suitable for mammography with synchrotron radiation in order to test the validity of the novel algorithm in extracting values of linear attenuation coefficients integrated over the sample thickness. It is demonstrated that the quantitative data are in good agreement with the theoretical values of linear attenuation coefficients calculated on the hypothesis of the breast with a given composition. The results are promising and encourage the current efforts to apply the method in mammography with synchrotron radiation.
2015
edge illumination; mammography; phase contrast; phase retrieval; synchrotron radiation; Light; Mammography; Radiation Dosage; Synchrotrons; X-Rays; Algorithms; Radiological and Ultrasound Technology; Radiology, Nuclear Medicine and Imaging
01 Pubblicazione su rivista::01a Articolo in rivista
A simplified edge illumination set-up for quantitative phase contrast mammography with synchrotron radiation at clinical doses / Longo, Mariaconcetta; Rigon, Luigi; Lopez, Frances C. M.; Chen, Rongchang; Dreossi, Diego; Zanconati, Fabrizio; Longo, Renata. - In: PHYSICS IN MEDICINE AND BIOLOGY. - ISSN 0031-9155. - ELETTRONICO. - 60:3(2015), pp. N21-N34. [10.1088/0031-9155/60/3/N21]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1071822
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 8
social impact