A series of small size 22 Na point sources have been simulated using GATE, following the NEMA protocol, and applied to the geometry of a brain dedicated PET so-called MindView. The simulated scanner geometry is currently composed by one ring of 20 detector blocks with a distance of about 330 mm between opposite detectors. Each detector module includes a monolithic LYSO crystal of 50 × 50 × 20 mm 3 and a custom 12 × 12 SiPM array (TSV-type, SensL). The system defines an axial and trans-axial fields of view (FoV) of about 46 mm and 240 mm, respectively. The scintillation light inside the monolithic block was simulated in GEANT4 and is sampled following the readout scheme providing information on each SiPM array row and column (12 + 12). Out of this, three different approaches for the determination of the impact coordinates within the monolithic crystals have been considered. A first approach makes use of the center of gravity (CoG) calculation for the planar XY coordinates but the photon impact depth of interaction (DoI) is not considered. The other methods obtain the XY coordinates using the Raise to Power algorithm (RTP) before CoG calculation, with both discrete (second method) and continuous (third one) DoI information. The data is introduced on the STIR platform for reconstruction using a filtered back projection 3D re-projected algorithm (FBP3DRP). Spatial resolutions ranging in the intervals [2 mm, 3.4 mm], [2.3 mm, 3.3mm] and [2.2 mm, 2.3 mm] have been obtained for the radial, tangential and axial directions, respectively. The points sources were positioned at a radial offset of 100 mm radially out of the center.

Performance evaluation of the MindView PET using GATE and STIR / Sanchez, Sebastian; Preziosi, Enrico; González, Antonio J.; Conde, Pablo; Iborra, Amadeo; Moliner, Laura; Borrazzo, Cristian; Aguilar, Albert; Bellido, Pablo; Díaz-Caballero, Elena; García-Garrigos, Juan J.; Grau-Ruiz, Daniel; González-Montoro, Andrea; Rigla, Juan P.; Rodríguez-Álvarez, María J.; Seimetz, Michael; Sánchez, Filomeno; Bettiol, Marco; Soriano, Antonio; Pani, Roberto; Correcher, Carlos; Benlloch, José M.. - 2017:(2017), pp. 1-5. (Intervento presentato al convegno 2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop, NSS/MIC/RTSD 2016 tenutosi a Strasbourg, France nel 2016) [10.1109/NSSMIC.2016.8069533].

Performance evaluation of the MindView PET using GATE and STIR

Preziosi, Enrico;Borrazzo, Cristian;Bettiol, Marco;Pani, Roberto;
2017

Abstract

A series of small size 22 Na point sources have been simulated using GATE, following the NEMA protocol, and applied to the geometry of a brain dedicated PET so-called MindView. The simulated scanner geometry is currently composed by one ring of 20 detector blocks with a distance of about 330 mm between opposite detectors. Each detector module includes a monolithic LYSO crystal of 50 × 50 × 20 mm 3 and a custom 12 × 12 SiPM array (TSV-type, SensL). The system defines an axial and trans-axial fields of view (FoV) of about 46 mm and 240 mm, respectively. The scintillation light inside the monolithic block was simulated in GEANT4 and is sampled following the readout scheme providing information on each SiPM array row and column (12 + 12). Out of this, three different approaches for the determination of the impact coordinates within the monolithic crystals have been considered. A first approach makes use of the center of gravity (CoG) calculation for the planar XY coordinates but the photon impact depth of interaction (DoI) is not considered. The other methods obtain the XY coordinates using the Raise to Power algorithm (RTP) before CoG calculation, with both discrete (second method) and continuous (third one) DoI information. The data is introduced on the STIR platform for reconstruction using a filtered back projection 3D re-projected algorithm (FBP3DRP). Spatial resolutions ranging in the intervals [2 mm, 3.4 mm], [2.3 mm, 3.3mm] and [2.2 mm, 2.3 mm] have been obtained for the radial, tangential and axial directions, respectively. The points sources were positioned at a radial offset of 100 mm radially out of the center.
2017
2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop, NSS/MIC/RTSD 2016
radiology, nuclear medicine and imaging; instrumentation; nuclear and high energy physics; electronic, optical and magnetic materials
04 Pubblicazione in atti di convegno::04b Atto di convegno in volume
Performance evaluation of the MindView PET using GATE and STIR / Sanchez, Sebastian; Preziosi, Enrico; González, Antonio J.; Conde, Pablo; Iborra, Amadeo; Moliner, Laura; Borrazzo, Cristian; Aguilar, Albert; Bellido, Pablo; Díaz-Caballero, Elena; García-Garrigos, Juan J.; Grau-Ruiz, Daniel; González-Montoro, Andrea; Rigla, Juan P.; Rodríguez-Álvarez, María J.; Seimetz, Michael; Sánchez, Filomeno; Bettiol, Marco; Soriano, Antonio; Pani, Roberto; Correcher, Carlos; Benlloch, José M.. - 2017:(2017), pp. 1-5. (Intervento presentato al convegno 2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop, NSS/MIC/RTSD 2016 tenutosi a Strasbourg, France nel 2016) [10.1109/NSSMIC.2016.8069533].
File allegati a questo prodotto
File Dimensione Formato  
Sanchez_Performance-evaluation_2016.pdf

solo gestori archivio

Tipologia: Versione editoriale (versione pubblicata con il layout dell'editore)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 528.52 kB
Formato Adobe PDF
528.52 kB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1071461
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 0
social impact