The Sprint FidelisTM ICD lead has a higher than expected incidence of conductor fractures. Their identification is important to prevent inappropriate shocks. The Sensing Integrity CounterTM (SIC), which stores the cumulative number of nonphysiologically short ventricular intervals ≤ 130 ms, is a diagnostic feature designed to enhance early identification of lead failures. Values greater than 300 are abnormal.
High sensing integrity counter in a patient with a Sprint Fidelis defibrillation lead: what is the cause? / Nguyen, BICH LIEN; Swerdlow, C; Gaudio, Carlo; Gang, Es. - In: HEART RHYTHM. - ISSN 1547-5271. - STAMPA. - 6(4):6(2009), pp. 579-581. [10.1016/j.hrthm.2008.11.021]
High sensing integrity counter in a patient with a Sprint Fidelis defibrillation lead: what is the cause?
NGUYEN, BICH LIEN;GAUDIO, Carlo;
2009
Abstract
The Sprint FidelisTM ICD lead has a higher than expected incidence of conductor fractures. Their identification is important to prevent inappropriate shocks. The Sensing Integrity CounterTM (SIC), which stores the cumulative number of nonphysiologically short ventricular intervals ≤ 130 ms, is a diagnostic feature designed to enhance early identification of lead failures. Values greater than 300 are abnormal.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.