Background The importance of measuring aortic pulse wave velocity (PWVao), aortic augmentation index (Aix) and central systolic blood pressure (SBPao) has been shown under different clinical conditions; however, information on these parameters is hard to obtain. The aim of this study was to evaluate the accuracy of a new, easily applicable oscillometric device (Arteriograph), determining these parameters simultaneously, against invasive measurements. Methods Aortic Aix, SBPao and PWVao were measured invasively during cardiac catheterization in 16, 55 and 22 cases, respectively, and compared with the values measured by the Arteriograph. Results We found strong correlation between the invasively measured aortic Aix and the oscillometrically measured brachial Aix on either beat-to-beat or mean value per patient basis (r=0.9, P<0.001; r=0.94, P<0.001), which allowed the noninvasive calculation of the aortic Aix without using generalized transfer function. Similarly strong correlation (r=0.95, P<0.001) was found between the invasively measured and the noninvasively calculated central SBPao; furthermore, the BHS assessment of the paired differences fulfilled the 'B' grading. The PWVao values measured invasively and by Arteriograph were 9.41 +/- 1.8 m/s and 9.46 +/- 1.8 m/s, respectively (mean +/- SD); furthermore, the Pearson's correlation was 0.91 (P<0.001). The limits of agreement were 11.4% for aortic Aix and 1.59 m/s for PWVao. Conclusion Aix, SBPao and PWVao, measured oscillometrically, showed strong correlation with the invasively obtained values. The observed limits of agreement are encouragingly low for accepting the method for clinical use. Our results suggest that the PWVao values, measured by Arteriograph, are close to the true aortic PWV, determined invasively. J Hypertens 28: 2068-2075 (C) 2010 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.

Invasive validation of a new oscillometric device (Arteriograph) for measuring augmentation index, central blood pressure and aortic pulse wave velocity / Ivan G., Horvath; Adam, Nemeth; Zsofia, Lenkey; Alessandri, Nicola; Fabrizio, Tufano; Pál, Kis; Balázs, Gaszner; Attila, Cziraki. - In: JOURNAL OF HYPERTENSION. - ISSN 0263-6352. - 28:10(2010), pp. 2068-2075. [10.1097/hjh.0b013e32833c8a1a]

Invasive validation of a new oscillometric device (Arteriograph) for measuring augmentation index, central blood pressure and aortic pulse wave velocity

ALESSANDRI, Nicola;
2010

Abstract

Background The importance of measuring aortic pulse wave velocity (PWVao), aortic augmentation index (Aix) and central systolic blood pressure (SBPao) has been shown under different clinical conditions; however, information on these parameters is hard to obtain. The aim of this study was to evaluate the accuracy of a new, easily applicable oscillometric device (Arteriograph), determining these parameters simultaneously, against invasive measurements. Methods Aortic Aix, SBPao and PWVao were measured invasively during cardiac catheterization in 16, 55 and 22 cases, respectively, and compared with the values measured by the Arteriograph. Results We found strong correlation between the invasively measured aortic Aix and the oscillometrically measured brachial Aix on either beat-to-beat or mean value per patient basis (r=0.9, P<0.001; r=0.94, P<0.001), which allowed the noninvasive calculation of the aortic Aix without using generalized transfer function. Similarly strong correlation (r=0.95, P<0.001) was found between the invasively measured and the noninvasively calculated central SBPao; furthermore, the BHS assessment of the paired differences fulfilled the 'B' grading. The PWVao values measured invasively and by Arteriograph were 9.41 +/- 1.8 m/s and 9.46 +/- 1.8 m/s, respectively (mean +/- SD); furthermore, the Pearson's correlation was 0.91 (P<0.001). The limits of agreement were 11.4% for aortic Aix and 1.59 m/s for PWVao. Conclusion Aix, SBPao and PWVao, measured oscillometrically, showed strong correlation with the invasively obtained values. The observed limits of agreement are encouragingly low for accepting the method for clinical use. Our results suggest that the PWVao values, measured by Arteriograph, are close to the true aortic PWV, determined invasively. J Hypertens 28: 2068-2075 (C) 2010 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
2010
arterial stiffness; arteriograph; augmentation index; invasive measurement; oscillometric method; pulse wave velocity; validation
01 Pubblicazione su rivista::01a Articolo in rivista
Invasive validation of a new oscillometric device (Arteriograph) for measuring augmentation index, central blood pressure and aortic pulse wave velocity / Ivan G., Horvath; Adam, Nemeth; Zsofia, Lenkey; Alessandri, Nicola; Fabrizio, Tufano; Pál, Kis; Balázs, Gaszner; Attila, Cziraki. - In: JOURNAL OF HYPERTENSION. - ISSN 0263-6352. - 28:10(2010), pp. 2068-2075. [10.1097/hjh.0b013e32833c8a1a]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/104079
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 81
  • Scopus 313
  • ???jsp.display-item.citation.isi??? 304
social impact