Purpose: To evaluate the accuracy and repeatability of a computer-generated Pelli–Robson test displayed on liquid crystal display (LCD) systems compared to a standard Pelli–Robson chart. Methods: Two different randomized crossover experiments were carried out for two different LCD systems for 32 subjects: 6 females and 10 males (40.5 ± 13.0 years) and 9 females and 7 males (27.8 ± 12.2 years), respectively, in the first and second experiment. Two repeated measurements were taken with the printed Pelli–Robson test and with the LCDs at 1 and 3 m. To test LCD reliability, measurements were repeated after 1 week. Results: In Experiment 1, contrast sensitivity (CS) measured with LCD1 resulted significantly higher than Pelli–Robson both at 1 and at 3 m of about 0.20 log 1/C in both eyes (p < 0.01). Bland–Altman plots showed a proportional bias for LCD1 measures. LCD1 measurements showed reasonable repeatability: ICC was 0.83 and 0.65 at 1 and 3 m, respectively. In Experiment 2, CS measured with LCD2 resulted significantly lower than Pelli–Robson both at 1 and at 3 m of about 0.10 log 1/C in both eyes (p < 0.01). Bland–Altman plots did not show any proportional bias for LCD2 measures. LCD2 measurements showed sufficient repeatability: ICC resulted 0.51 and 0.65 at 1 and 3 m, respectively. Conclusions: Computer-generated versions of Pelli–Robson test, displayed on LCD systems, do not provide accurate results compared to classic Pelli–Robson printed version. Clinicians should consider that Pelli–Robson computer-generated versions could be non-interchangeable to the printed version

How accurate is an LCD screen version of the Pelli/Robson test? / Zeri, Fabrizio; Calcatelli, Paolo; Funaro, Eleonora; Martelli, Marialuisa; Naroo, Shehzad A.. - In: INTERNATIONAL OPHTHALMOLOGY. - ISSN 0165-5701. - STAMPA. - 38(2018), pp. 1473-1484. [10.1007/s10792-017-0609-0]

How accurate is an LCD screen version of the Pelli/Robson test?

MARTELLI, Marialuisa;
2018

Abstract

Purpose: To evaluate the accuracy and repeatability of a computer-generated Pelli–Robson test displayed on liquid crystal display (LCD) systems compared to a standard Pelli–Robson chart. Methods: Two different randomized crossover experiments were carried out for two different LCD systems for 32 subjects: 6 females and 10 males (40.5 ± 13.0 years) and 9 females and 7 males (27.8 ± 12.2 years), respectively, in the first and second experiment. Two repeated measurements were taken with the printed Pelli–Robson test and with the LCDs at 1 and 3 m. To test LCD reliability, measurements were repeated after 1 week. Results: In Experiment 1, contrast sensitivity (CS) measured with LCD1 resulted significantly higher than Pelli–Robson both at 1 and at 3 m of about 0.20 log 1/C in both eyes (p < 0.01). Bland–Altman plots showed a proportional bias for LCD1 measures. LCD1 measurements showed reasonable repeatability: ICC was 0.83 and 0.65 at 1 and 3 m, respectively. In Experiment 2, CS measured with LCD2 resulted significantly lower than Pelli–Robson both at 1 and at 3 m of about 0.10 log 1/C in both eyes (p < 0.01). Bland–Altman plots did not show any proportional bias for LCD2 measures. LCD2 measurements showed sufficient repeatability: ICC resulted 0.51 and 0.65 at 1 and 3 m, respectively. Conclusions: Computer-generated versions of Pelli–Robson test, displayed on LCD systems, do not provide accurate results compared to classic Pelli–Robson printed version. Clinicians should consider that Pelli–Robson computer-generated versions could be non-interchangeable to the printed version
2018
computerized-based stimuli; contrast sensitivity; LCD systems; Pelli-Robson test; ophthalmology
01 Pubblicazione su rivista::01a Articolo in rivista
How accurate is an LCD screen version of the Pelli/Robson test? / Zeri, Fabrizio; Calcatelli, Paolo; Funaro, Eleonora; Martelli, Marialuisa; Naroo, Shehzad A.. - In: INTERNATIONAL OPHTHALMOLOGY. - ISSN 0165-5701. - STAMPA. - 38(2018), pp. 1473-1484. [10.1007/s10792-017-0609-0]
File allegati a questo prodotto
File Dimensione Formato  
Zeri_How_2017.pdf

solo utenti autorizzati

Note: paper
Tipologia: Documento in Post-print (versione successiva alla peer review e accettata per la pubblicazione)
Licenza: Tutti i diritti riservati (All rights reserved)
Dimensione 450.02 kB
Formato Adobe PDF
450.02 kB Adobe PDF   Contatta l'autore

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/1010935
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 6
social impact