Doppler ultrasound (US) is a valuable tool to measure blood flow in the transplanted kidney, but its operator-dependence can greatly affect repeatability and reproducibility of measurements. Aim of this work was to evaluate intraobserver and interobserver variability in measuring the resistive index (RI) in renal transplants. Ten renal transplant recipients were randomly selected among those undergoing follow-up and examined by two operators (FG and LB) with 3.5 MHz and 10 MHz scanheads to assess the variability of RI measurements. Each observer obtained two measurements of the RI with each scanhead within a 10-15 minutes' period. In all, 80 measurements were made, 4 per patient per observer. The statistical analysis included two-tailed Student's t-test for paired data and calculation of repeatability/reproducibility coefficients. Student's t-test analysis demonstrated a statistically significant difference (p = 0.037) between the means of the first and second measurements by FG with the 3.5 MHz scanhead and the first and the second measurements by LB with the same scanhead. Differences between the other means were not statistically significant. Intraobserver variability ranged 0.03 units (or 2.07%) and 0.07 units (or 4.24%), while interobserver variability was 0.04 units with both 3.5 and 10 MHz scanheads, or 3.61 and 3.73%, respectively. Doppler US of renal transplants has statistically quantifiable operator-dependent variability: the possible evidence of statistically significant differences can be minimized by having the same operator make the measurements. However, RI variations ranging 0.02 to 0.04 units should not be considered significant.

[A statistical evaluation of the variability in the measurements of the resistive index in kidney transplantation] / Giovagnorio, Francesco; Ballesio, Laura. - In: LA RADIOLOGIA MEDICA. - ISSN 0033-8362. - STAMPA. - 99:3(2000), pp. 174-176.

[A statistical evaluation of the variability in the measurements of the resistive index in kidney transplantation].

GIOVAGNORIO, Francesco;BALLESIO, Laura
2000

Abstract

Doppler ultrasound (US) is a valuable tool to measure blood flow in the transplanted kidney, but its operator-dependence can greatly affect repeatability and reproducibility of measurements. Aim of this work was to evaluate intraobserver and interobserver variability in measuring the resistive index (RI) in renal transplants. Ten renal transplant recipients were randomly selected among those undergoing follow-up and examined by two operators (FG and LB) with 3.5 MHz and 10 MHz scanheads to assess the variability of RI measurements. Each observer obtained two measurements of the RI with each scanhead within a 10-15 minutes' period. In all, 80 measurements were made, 4 per patient per observer. The statistical analysis included two-tailed Student's t-test for paired data and calculation of repeatability/reproducibility coefficients. Student's t-test analysis demonstrated a statistically significant difference (p = 0.037) between the means of the first and second measurements by FG with the 3.5 MHz scanhead and the first and the second measurements by LB with the same scanhead. Differences between the other means were not statistically significant. Intraobserver variability ranged 0.03 units (or 2.07%) and 0.07 units (or 4.24%), while interobserver variability was 0.04 units with both 3.5 and 10 MHz scanheads, or 3.61 and 3.73%, respectively. Doppler US of renal transplants has statistically quantifiable operator-dependent variability: the possible evidence of statistically significant differences can be minimized by having the same operator make the measurements. However, RI variations ranging 0.02 to 0.04 units should not be considered significant.
2000
article; doppler; doppler flowmetry; echography; emtree medical terms: adult; female; follow up; follow-up studies; human; humans; instrumentation; kidney circulation; kidney transplantation; male; middle aged; observer variation; physiology; random allocation; randomization; renal circulation; reproducibility; reproducibility of results; statistics; ultrasonography; vascular resistance; vascular resistance mesh: adult
01 Pubblicazione su rivista::01a Articolo in rivista
[A statistical evaluation of the variability in the measurements of the resistive index in kidney transplantation] / Giovagnorio, Francesco; Ballesio, Laura. - In: LA RADIOLOGIA MEDICA. - ISSN 0033-8362. - STAMPA. - 99:3(2000), pp. 174-176.
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/67538
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 4
  • ???jsp.display-item.citation.isi??? ND
social impact