Introduction: The purpose of this study was to evaluate the effect of autoclave sterilization on cyclic fatigue resistance of rotary endodontic instruments made of traditional and new nickel-titanium (NiTi) alloys. Methods: Four NiTi rotary endodontic instruments of the same size (tip diameter 0.40 mm and constant .04 taper) were selected: K3, Mtwo, Vortex, and K3 XF prototypes. Each group was then divided into 2 subgroups, unsterilized instruments and sterilized instruments. The sterilized instruments were subjected to 10 cycles of autoclave sterilization. Twelve files from each different subgroup were tested for cyclic fatigue resistance. Means and standard deviations of number of cycles to failure (NCF) and fragment length of the fractured tip were calculated for each group, and data were statistically analyzed (P < .05). Results: Comparing the results between unsterilized and sterilized instruments for each type of file, differences were statistically significant (P < .05) only between sterilized and unsterilized K3XF files (762 versus 651 NCF). The other instruments did not show significant differences (P > .05) in the mean NCF as a result of sterilization cycles (K3, 424 versus 439 NCF; Mtwo, 409 versus 419 NCF; Vortex, 454 versus 480 NCF). Comparing the results among the different groups, K3 XF (either sterilized or not) showed a mean NCF significantly higher than all other files(P < .05). Conclusions: Repeated cycles of autoclave sterilization do not seem to influence the mechanical properties of NiTi endodontic instruments except for the K3 XF prototypes of rotary instruments that demonstrated a significant increase of cyclic fatigue resistance. (J Endod 2012;38:222-225)

Experimental Evaluation on the Influence of Autoclave Sterilization on the Cyclic Fatigue of New Nickel-Titanium Rotary Instruments / Gianluca, Plotino; Alberto, Costanzo; Grande, NICOLA MARIA; Renata, Petrovic; Testarelli, Luca; Gambarini, Gianluca. - In: JOURNAL OF ENDODONTICS. - ISSN 0099-2399. - 38:2(2012), pp. 222-225. [10.1016/j.joen.2011.10.017]

Experimental Evaluation on the Influence of Autoclave Sterilization on the Cyclic Fatigue of New Nickel-Titanium Rotary Instruments

GRANDE, NICOLA MARIA;TESTARELLI, Luca;GAMBARINI, Gianluca
2012

Abstract

Introduction: The purpose of this study was to evaluate the effect of autoclave sterilization on cyclic fatigue resistance of rotary endodontic instruments made of traditional and new nickel-titanium (NiTi) alloys. Methods: Four NiTi rotary endodontic instruments of the same size (tip diameter 0.40 mm and constant .04 taper) were selected: K3, Mtwo, Vortex, and K3 XF prototypes. Each group was then divided into 2 subgroups, unsterilized instruments and sterilized instruments. The sterilized instruments were subjected to 10 cycles of autoclave sterilization. Twelve files from each different subgroup were tested for cyclic fatigue resistance. Means and standard deviations of number of cycles to failure (NCF) and fragment length of the fractured tip were calculated for each group, and data were statistically analyzed (P < .05). Results: Comparing the results between unsterilized and sterilized instruments for each type of file, differences were statistically significant (P < .05) only between sterilized and unsterilized K3XF files (762 versus 651 NCF). The other instruments did not show significant differences (P > .05) in the mean NCF as a result of sterilization cycles (K3, 424 versus 439 NCF; Mtwo, 409 versus 419 NCF; Vortex, 454 versus 480 NCF). Comparing the results among the different groups, K3 XF (either sterilized or not) showed a mean NCF significantly higher than all other files(P < .05). Conclusions: Repeated cycles of autoclave sterilization do not seem to influence the mechanical properties of NiTi endodontic instruments except for the K3 XF prototypes of rotary instruments that demonstrated a significant increase of cyclic fatigue resistance. (J Endod 2012;38:222-225)
2012
autoclave sterilization; cyclic fatigue; niti instruments
01 Pubblicazione su rivista::01a Articolo in rivista
Experimental Evaluation on the Influence of Autoclave Sterilization on the Cyclic Fatigue of New Nickel-Titanium Rotary Instruments / Gianluca, Plotino; Alberto, Costanzo; Grande, NICOLA MARIA; Renata, Petrovic; Testarelli, Luca; Gambarini, Gianluca. - In: JOURNAL OF ENDODONTICS. - ISSN 0099-2399. - 38:2(2012), pp. 222-225. [10.1016/j.joen.2011.10.017]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/431919
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 21
  • Scopus 74
  • ???jsp.display-item.citation.isi??? 69
social impact