A rare case of cochlear implant surgery complication is described: air collection behind the ear. a 61-year-old male with a 20-year history of progressive bilateral profound sensorineural hearing loss underwent cochlear implant surgery on the left ear with Clarion Hi-Res 90K (Advanced Bionics, Sylmar, CA, USA). Ten days after surgery, the patient presented visible tumefaction behind the ear corresponding to the receiver-stimulator. ultrasonography, with a probe of 30 MHz, of the surface behind the ear showed formation of probable liquid content but aspiration with a 20-gauge needle yielded 30 cm(3) of air without blood or pus. An unusual case is described of a minor complication occurring in an adult cochlear implant patient presenting a swelling behind the ear that was found to be collected air. Although ultrasonography can be useful to evaluate localization of swelling behind the ear and to differentiate between liquid and solid collection, it is not useful for identification of air collection. The Eustachian tube test can show an abnormally patent tube and prevent this complication.

Air-collection behind the ear: a rare complication in cochlear implant surgery / Viccaro, Marika; DE SETA, Elio; Covelli, Edoardo; Balsamo, Giorgio; Filipo, Roberto. - In: ACTA OTORHINOLARYNGOLOGICA ITALICA. - ISSN 0392-100X. - 28:4(2008), pp. 218-220.

Air-collection behind the ear: a rare complication in cochlear implant surgery

VICCARO, MARIKA;DE SETA, Elio;COVELLI, EDOARDO;BALSAMO, GIORGIO;FILIPO, Roberto
2008

Abstract

A rare case of cochlear implant surgery complication is described: air collection behind the ear. a 61-year-old male with a 20-year history of progressive bilateral profound sensorineural hearing loss underwent cochlear implant surgery on the left ear with Clarion Hi-Res 90K (Advanced Bionics, Sylmar, CA, USA). Ten days after surgery, the patient presented visible tumefaction behind the ear corresponding to the receiver-stimulator. ultrasonography, with a probe of 30 MHz, of the surface behind the ear showed formation of probable liquid content but aspiration with a 20-gauge needle yielded 30 cm(3) of air without blood or pus. An unusual case is described of a minor complication occurring in an adult cochlear implant patient presenting a swelling behind the ear that was found to be collected air. Although ultrasonography can be useful to evaluate localization of swelling behind the ear and to differentiate between liquid and solid collection, it is not useful for identification of air collection. The Eustachian tube test can show an abnormally patent tube and prevent this complication.
2008
air collection; cochlear implant; minor complications; surgery
01 Pubblicazione su rivista::01a Articolo in rivista
Air-collection behind the ear: a rare complication in cochlear implant surgery / Viccaro, Marika; DE SETA, Elio; Covelli, Edoardo; Balsamo, Giorgio; Filipo, Roberto. - In: ACTA OTORHINOLARYNGOLOGICA ITALICA. - ISSN 0392-100X. - 28:4(2008), pp. 218-220.
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/40441
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 6
social impact