In this work, we report and compare the results of gamma-ray spectroscopy measurements performed using a LaBr3:Ce crystal coupled to a photornultiplier tube with modified base and to a Si-avalanche photodiode readout by standard electronics. Energy resolution values have been obtained in the range from 59.5 keV (Am-241) to 1332 keV (Co-60). Good linearity in energy response has been found. Energy resolution values obtained with avalanche photodiode readout are underestimated due to the small active area of the presently available photosensor with respect to crystal dimension. (c) 2006 Elsevier B.V. All rights reserved.

Si-APD readout for LaBr3 : Ce scintillator / Raffaele, Scafe; Pani, Roberto; Pellegrini, Rosanna; Giorgia, Iurlaro; Livia, Montani; Cinti, Maria Nerina. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - STAMPA. - 571:1-2 SPEC. ISS.(2007), pp. 355-357. (Intervento presentato al convegno 1st International Conference on Molecular Imaging Technology tenutosi a Marseille, FRANCE nel MAY 09-12, 2006) [10.1016/j.nima.2006.10.108].

Si-APD readout for LaBr3 : Ce scintillator

PANI, Roberto;PELLEGRINI, Rosanna;CINTI, Maria Nerina
2007

Abstract

In this work, we report and compare the results of gamma-ray spectroscopy measurements performed using a LaBr3:Ce crystal coupled to a photornultiplier tube with modified base and to a Si-avalanche photodiode readout by standard electronics. Energy resolution values have been obtained in the range from 59.5 keV (Am-241) to 1332 keV (Co-60). Good linearity in energy response has been found. Energy resolution values obtained with avalanche photodiode readout are underestimated due to the small active area of the presently available photosensor with respect to crystal dimension. (c) 2006 Elsevier B.V. All rights reserved.
2007
gamma-ray spectroscopy; labr3 : ce; labr3:ce; nuclear medicine; scintillator; γ-ray spectroscopy
01 Pubblicazione su rivista::01a Articolo in rivista
Si-APD readout for LaBr3 : Ce scintillator / Raffaele, Scafe; Pani, Roberto; Pellegrini, Rosanna; Giorgia, Iurlaro; Livia, Montani; Cinti, Maria Nerina. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - STAMPA. - 571:1-2 SPEC. ISS.(2007), pp. 355-357. (Intervento presentato al convegno 1st International Conference on Molecular Imaging Technology tenutosi a Marseille, FRANCE nel MAY 09-12, 2006) [10.1016/j.nima.2006.10.108].
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/365741
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 4
social impact