This paper characterizes the moment measurement accuracy for a novel parallel spherical robot (SR) for dynamic posturography, controllable by position or impedance. The SR consists of three linearmotors placed on a support base, amoving base, and three passive arms equipped with uniaxial load cells permitting impedance controlled perturbations. To evaluate the accuracy, a subject stood still on the SR, set in position control mode, while selected sinusoidal trajectories were applied. The moments computed by the load cells were compared to the value measured by a six-component force platform, placed on top of the rotating base. For the intended application of the SR, the errors were negligible with the worse case of only 4 Nm in a total of 15 trials (five conditions, three repetitions). The observed moment error was related mainly to the intrinsic accuracy of the sensors, equal to about 7 N. To demonstrate clinical applicability, the platform was set to impedance control mode and a protocol was tested with a 12-year-old girl with brain injury and a group of four healthy subjects. In total, 24 trials (eight conditions, three repetitions) were recorded for each subject. The results of this pilot study identified distinctive postural behaviors and therefore showed that the SR can be considered as an effective tool for dynamic posturography.

Moment Measurement Accuracy of a Parallel Spherical Robot for Dynamic Posturography / Cappa, Paolo; Jackson, Jennifer; Patane', Fabrizio. - In: IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING. - ISSN 0018-9294. - STAMPA. - 57:5(2010), pp. 1198-1208. [10.1109/tbme.2009.2037807]

Moment Measurement Accuracy of a Parallel Spherical Robot for Dynamic Posturography

CAPPA, Paolo;JACKSON, JENNIFER;PATANE', FABRIZIO
2010

Abstract

This paper characterizes the moment measurement accuracy for a novel parallel spherical robot (SR) for dynamic posturography, controllable by position or impedance. The SR consists of three linearmotors placed on a support base, amoving base, and three passive arms equipped with uniaxial load cells permitting impedance controlled perturbations. To evaluate the accuracy, a subject stood still on the SR, set in position control mode, while selected sinusoidal trajectories were applied. The moments computed by the load cells were compared to the value measured by a six-component force platform, placed on top of the rotating base. For the intended application of the SR, the errors were negligible with the worse case of only 4 Nm in a total of 15 trials (five conditions, three repetitions). The observed moment error was related mainly to the intrinsic accuracy of the sensors, equal to about 7 N. To demonstrate clinical applicability, the platform was set to impedance control mode and a protocol was tested with a 12-year-old girl with brain injury and a group of four healthy subjects. In total, 24 trials (eight conditions, three repetitions) were recorded for each subject. The results of this pilot study identified distinctive postural behaviors and therefore showed that the SR can be considered as an effective tool for dynamic posturography.
2010
dynamic posturography; parallel robot; postural control and balance; rotating platform; spherical robot; spherical robot (sr)
01 Pubblicazione su rivista::01a Articolo in rivista
Moment Measurement Accuracy of a Parallel Spherical Robot for Dynamic Posturography / Cappa, Paolo; Jackson, Jennifer; Patane', Fabrizio. - In: IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING. - ISSN 0018-9294. - STAMPA. - 57:5(2010), pp. 1198-1208. [10.1109/tbme.2009.2037807]
File allegati a questo prodotto
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11573/361413
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 15
  • ???jsp.display-item.citation.isi??? 11
social impact